Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Radiochromic film based transit dosimetry for verification of dose delivery with intensity modulated radiotherapy

Full metadata record
DC Field Value Language
dc.contributor.authorChung, Kwangzoo-
dc.contributor.authorYoon, Myonggeun-
dc.contributor.authorSon, Jaeman-
dc.contributor.authorPark, Sung Yong-
dc.contributor.authorLee, Kiho-
dc.contributor.authorShin, Dongho-
dc.contributor.authorLim, Young Kyung-
dc.contributor.authorLee, Se Byeong-
dc.date.accessioned2021-09-06T04:51:37Z-
dc.date.available2021-09-06T04:51:37Z-
dc.date.created2021-06-14-
dc.date.issued2013-02-
dc.identifier.issn0094-2405-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/104114-
dc.description.abstractPurpose: To evaluate the transit dose based patient specific quality assurance (QA) of intensity modulated radiation therapy (IMRT) for verification of the accuracy of dose delivered to the patient. Methods: Five IMRT plans were selected and utilized to irradiate a homogeneous plastic water phantom and an inhomogeneous anthropomorphic phantom. The transit dose distribution was measured with radiochromic film and was compared with the computed dose map on the same plane using a gamma index with a 3% dose and a 3 mm distance-to-dose agreement tolerance limit. Results: While the average gamma index for comparisons of dose distributions was less than one for 98.9% of all pixels from the transit dose with the homogeneous phantom, the passing rate was reduced to 95.0% for the transit dose with the inhomogeneous phantom. Transit doses due to a 5 mm setup error may cause up to a 50% failure rate of the gamma index. Conclusions: Transit dose based IMRT QA may be superior to the traditional QA method since the former can show whether the inhomogeneity correction algorithm from TPS is accurate. In addition, transit dose based IMRT QA can be used to verify the accuracy of the dose delivered to the patient during treatment by revealing significant increases in the failure rate of the gamma index resulting from errors in patient positioning during treatment. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4788658]-
dc.languageEnglish-
dc.language.isoen-
dc.publisherAMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS-
dc.subjectRADIATION-THERAPY-
dc.subjectQUALITY-ASSURANCE-
dc.subjectPROSTATE-CANCER-
dc.subjectIMRT-
dc.subjectACCURACY-
dc.subjectHEAD-
dc.subjectNECK-
dc.subjectDISTRIBUTIONS-
dc.subjectMARKERS-
dc.subjectMOTION-
dc.titleRadiochromic film based transit dosimetry for verification of dose delivery with intensity modulated radiotherapy-
dc.typeArticle-
dc.contributor.affiliatedAuthorYoon, Myonggeun-
dc.identifier.doi10.1118/1.4788658-
dc.identifier.scopusid2-s2.0-84873603069-
dc.identifier.wosid000314727700029-
dc.identifier.bibliographicCitationMEDICAL PHYSICS, v.40, no.2-
dc.relation.isPartOfMEDICAL PHYSICS-
dc.citation.titleMEDICAL PHYSICS-
dc.citation.volume40-
dc.citation.number2-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaRadiology, Nuclear Medicine & Medical Imaging-
dc.relation.journalWebOfScienceCategoryRadiology, Nuclear Medicine & Medical Imaging-
dc.subject.keywordPlusRADIATION-THERAPY-
dc.subject.keywordPlusQUALITY-ASSURANCE-
dc.subject.keywordPlusPROSTATE-CANCER-
dc.subject.keywordPlusIMRT-
dc.subject.keywordPlusACCURACY-
dc.subject.keywordPlusHEAD-
dc.subject.keywordPlusNECK-
dc.subject.keywordPlusDISTRIBUTIONS-
dc.subject.keywordPlusMARKERS-
dc.subject.keywordPlusMOTION-
dc.subject.keywordAuthorintensity modulated radiation therapy-
dc.subject.keywordAuthorquality assurance-
dc.subject.keywordAuthortransit dose-
dc.subject.keywordAuthorradiochromic film-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Myong geun photo

Yoon, Myong geun
보건과학대학 (바이오의공학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE