Interruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia -A case report-
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 조혜형 | - |
dc.contributor.author | 최성욱 | - |
dc.contributor.author | 박지용 | - |
dc.contributor.author | 신혜원 | - |
dc.date.accessioned | 2021-09-06T07:03:51Z | - |
dc.date.available | 2021-09-06T07:03:51Z | - |
dc.date.created | 2021-06-17 | - |
dc.date.issued | 2013 | - |
dc.identifier.issn | 1975-5171 | - |
dc.identifier.uri | https://scholar.korea.ac.kr/handle/2021.sw.korea/104857 | - |
dc.description.abstract | We report our experience of unexpected low expiratory concentration of sevoflurane 0.8 vol%, low inspired fraction of oxygen 31%,and high value of bispctral index (BIS) 62 with sevoflurlane vaporizer setting 2.0 vol% and the inspired fraction of oxygen 50% during general anesthesia of a 56-year-old man undergoing elective laparoscopic hemicolectomy. The anesthesia machine (Cato edition,Dräger, Lübeck, Germany) with a piston-based, electronically controlled ventilator was used for anesthesia. We checked thoroughly all the possible causes of the discrepancy of anesthetic gas concentration, and then detected the partial disconnection of fresh gas outlet. But there were no alarm signs of anesthetic monitor with adequate ventilator function. This malfunction of anesthetic gas supply without alarm signs may lead to hypoxia and awareness of patients. The intraoperative monitoring of anesthetic gas analyzer and BIS may be essential to detect the malfunction of anesthesia machine. | - |
dc.language | English | - |
dc.language.iso | en | - |
dc.publisher | 대한마취통증의학회 | - |
dc.title | Interruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia -A case report- | - |
dc.title.alternative | Interruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia -A case report- | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | 최성욱 | - |
dc.contributor.affiliatedAuthor | 신혜원 | - |
dc.identifier.bibliographicCitation | Anesthesia and Pain Medicine, v.8, no.4, pp.237 - 239 | - |
dc.relation.isPartOf | Anesthesia and Pain Medicine | - |
dc.citation.title | Anesthesia and Pain Medicine | - |
dc.citation.volume | 8 | - |
dc.citation.number | 4 | - |
dc.citation.startPage | 237 | - |
dc.citation.endPage | 239 | - |
dc.type.rims | ART | - |
dc.identifier.kciid | ART001814191 | - |
dc.description.journalClass | 2 | - |
dc.description.journalRegisteredClass | kci | - |
dc.description.journalRegisteredClass | other | - |
dc.subject.keywordAuthor | Alarm | - |
dc.subject.keywordAuthor | Disconnection | - |
dc.subject.keywordAuthor | Fresh gas outlet | - |
dc.subject.keywordAuthor | Ventilator. | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.