Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Interruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia -A case report-

Full metadata record
DC Field Value Language
dc.contributor.author조혜형-
dc.contributor.author최성욱-
dc.contributor.author박지용-
dc.contributor.author신혜원-
dc.date.accessioned2021-09-06T07:03:51Z-
dc.date.available2021-09-06T07:03:51Z-
dc.date.created2021-06-17-
dc.date.issued2013-
dc.identifier.issn1975-5171-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/104857-
dc.description.abstractWe report our experience of unexpected low expiratory concentration of sevoflurane 0.8 vol%, low inspired fraction of oxygen 31%,and high value of bispctral index (BIS) 62 with sevoflurlane vaporizer setting 2.0 vol% and the inspired fraction of oxygen 50% during general anesthesia of a 56-year-old man undergoing elective laparoscopic hemicolectomy. The anesthesia machine (Cato edition,Dräger, Lübeck, Germany) with a piston-based, electronically controlled ventilator was used for anesthesia. We checked thoroughly all the possible causes of the discrepancy of anesthetic gas concentration, and then detected the partial disconnection of fresh gas outlet. But there were no alarm signs of anesthetic monitor with adequate ventilator function. This malfunction of anesthetic gas supply without alarm signs may lead to hypoxia and awareness of patients. The intraoperative monitoring of anesthetic gas analyzer and BIS may be essential to detect the malfunction of anesthesia machine.-
dc.languageEnglish-
dc.language.isoen-
dc.publisher대한마취통증의학회-
dc.titleInterruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia -A case report--
dc.title.alternativeInterruption of fresh gas supply without alarm signs by the partial disconnection of fresh gas outlet during general anesthesia -A case report--
dc.typeArticle-
dc.contributor.affiliatedAuthor최성욱-
dc.contributor.affiliatedAuthor신혜원-
dc.identifier.bibliographicCitationAnesthesia and Pain Medicine, v.8, no.4, pp.237 - 239-
dc.relation.isPartOfAnesthesia and Pain Medicine-
dc.citation.titleAnesthesia and Pain Medicine-
dc.citation.volume8-
dc.citation.number4-
dc.citation.startPage237-
dc.citation.endPage239-
dc.type.rimsART-
dc.identifier.kciidART001814191-
dc.description.journalClass2-
dc.description.journalRegisteredClasskci-
dc.description.journalRegisteredClassother-
dc.subject.keywordAuthorAlarm-
dc.subject.keywordAuthorDisconnection-
dc.subject.keywordAuthorFresh gas outlet-
dc.subject.keywordAuthorVentilator.-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Shin, Hye Won photo

Shin, Hye Won
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE