Design and Testing of Augmented Reality-Based Fluorescence Imaging Goggle for Intraoperative Imaging-Guided Surgery
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Lee, Seung Hyun | - |
dc.contributor.author | Quan, Yu Hua | - |
dc.contributor.author | Kim, Min Sub | - |
dc.contributor.author | Kwon, Ki Hyeok | - |
dc.contributor.author | Choi, Byeong Hyeon | - |
dc.contributor.author | Kim, Hyun Koo | - |
dc.contributor.author | Kim, Beop-Min | - |
dc.date.accessioned | 2021-11-19T07:40:57Z | - |
dc.date.available | 2021-11-19T07:40:57Z | - |
dc.date.created | 2021-08-30 | - |
dc.date.issued | 2021-06 | - |
dc.identifier.issn | 2075-4418 | - |
dc.identifier.uri | https://scholar.korea.ac.kr/handle/2021.sw.korea/127971 | - |
dc.description.abstract | The different pathways between the position of a near-infrared camera and the user's eye limit the use of existing near-infrared fluorescence imaging systems for tumor margin assessments. By utilizing an optical system that precisely matches the near-infrared fluorescence image and the optical path of visible light, we developed an augmented reality (AR)-based fluorescence imaging system that provides users with a fluorescence image that matches the real-field, without requiring any additional algorithms. Commercial smart glasses, dichroic beam splitters, mirrors, and custom near-infrared cameras were employed to develop the proposed system, and each mount was designed and utilized. After its performance was assessed in the laboratory, preclinical experiments involving tumor detection and lung lobectomy in mice and rabbits by using indocyanine green (ICG) were conducted. The results showed that the proposed system provided a stable image of fluorescence that matched the actual site. In addition, preclinical experiments confirmed that the proposed system could be used to detect tumors using ICG and evaluate lung lobectomies. The AR-based intraoperative smart goggle system could detect fluorescence images for tumor margin assessments in animal models, without disrupting the surgical workflow in an operating room. Additionally, it was confirmed that, even when the system itself was distorted when worn, the fluorescence image consistently matched the actual site. | - |
dc.language | English | - |
dc.language.iso | en | - |
dc.publisher | MDPI | - |
dc.subject | NEAR-INFRARED FLUORESCENCE | - |
dc.subject | HUMAN SERUM-ALBUMIN | - |
dc.subject | INDOCYANINE GREEN | - |
dc.subject | NAVIGATION SYSTEM | - |
dc.subject | CANCER-SURGERY | - |
dc.subject | IDENTIFICATION | - |
dc.subject | RESECTION | - |
dc.subject | COLOR | - |
dc.title | Design and Testing of Augmented Reality-Based Fluorescence Imaging Goggle for Intraoperative Imaging-Guided Surgery | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | Kim, Hyun Koo | - |
dc.contributor.affiliatedAuthor | Kim, Beop-Min | - |
dc.identifier.doi | 10.3390/diagnostics11060927 | - |
dc.identifier.scopusid | 2-s2.0-85107433237 | - |
dc.identifier.wosid | 000667370900001 | - |
dc.identifier.bibliographicCitation | DIAGNOSTICS, v.11, no.6 | - |
dc.relation.isPartOf | DIAGNOSTICS | - |
dc.citation.title | DIAGNOSTICS | - |
dc.citation.volume | 11 | - |
dc.citation.number | 6 | - |
dc.type.rims | ART | - |
dc.type.docType | Article | - |
dc.description.journalClass | 1 | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | General & Internal Medicine | - |
dc.relation.journalWebOfScienceCategory | Medicine, General & Internal | - |
dc.subject.keywordPlus | NEAR-INFRARED FLUORESCENCE | - |
dc.subject.keywordPlus | HUMAN SERUM-ALBUMIN | - |
dc.subject.keywordPlus | INDOCYANINE GREEN | - |
dc.subject.keywordPlus | NAVIGATION SYSTEM | - |
dc.subject.keywordPlus | CANCER-SURGERY | - |
dc.subject.keywordPlus | IDENTIFICATION | - |
dc.subject.keywordPlus | RESECTION | - |
dc.subject.keywordPlus | COLOR | - |
dc.subject.keywordAuthor | augmented reality | - |
dc.subject.keywordAuthor | fluorescence image | - |
dc.subject.keywordAuthor | neuro-navigation | - |
dc.subject.keywordAuthor | smart goggle | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.