Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Pre-Prototype Stimulating and Recording Endotracheal Tube for Continuous Monitoring of the Recurrent Laryngeal Nerve During Thyroid Surgery

Authors
Zhang, DaqiWang, TieZhao, YishenSun, HuiPino, AntonellaKim, Hoon YubDionigi, Gianlorenzo
Issue Date
2021
Publisher
TAYLOR & FRANCIS INC
Keywords
pre-prototype; nerve monitoring; continuous nerve monitoring; recurrent laryngeal nerve; thyroid surgery; stimulation and recording endotracheal tube; EMG; amplitude; latency; waveform; morbidity
Citation
JOURNAL OF INVESTIGATIVE SURGERY, v.34, no.9, pp.1033 - 1043
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF INVESTIGATIVE SURGERY
Volume
34
Number
9
Start Page
1033
End Page
1043
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/130246
DOI
10.1080/08941939.2020.1734693
ISSN
0894-1939
Abstract
Introduction: Continuous intraoperative neural monitoring (C-IONM) is a promising technology used to decrease recurrent laryngeal nerve (RLN) damage during thyroid surgery. However, C-IONM use is limited due to its challenging application. C-IONM requires dissection of the carotid sheath and placement of an electrode around the vagus nerve (VN). In our study, we simultaneously stimulated and monitored the proximal RLN trans-tracheally using surface electrodes that were positioned solely on the endotracheal tube. Methods: We described the design, implementation, and testing of a pre-prototype, combined stimulating and recording endotracheal tube (SRET) that continually delivered current from tube edge, and tested the function of the RLN at the vocal cords (VC). The SRET was tested in vivo on 10 RLNs using 5 pigs as animal models. Results: The SRET was capable of the following, delivery of continuous trans-tracheal stimulation to the proximal RLN, continuous trans-tracheal stimulation-induced VC movement, and standard ipsilateral RLN biphasic waveforms with latency and amplitude; and recording evoked responses were recorded in the ipsilateral RLN. Conclusion: The pre-prototype SRET represents a possible advancement in technology because it simplified the C-IONM. Thus, the SRET provides a minimally invasive, alternative application to the C-IONM vagal nerve cuff electrodes.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Hoon Yub photo

Kim, Hoon Yub
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE