Performance evaluation of the Anysis-200 platelet function analyzer in cardiac patients

Full metadata record
DC Field Value Language
dc.contributor.authorPark, Seong Jun-
dc.contributor.authorYoon, Jung-
dc.contributor.authorSeo, Hong Seog-
dc.contributor.authorLim, Chae Seung-
dc.date.accessioned2022-02-23T05:40:39Z-
dc.date.available2022-02-23T05:40:39Z-
dc.date.created2022-02-21-
dc.date.issued2022-
dc.identifier.issn1386-0291-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/136585-
dc.description.abstractBACKGROUND: Platelet function analysis is crucial in assessing the hemostatic status to evaluate congenital and acquired platelet function defects. The Anysis-200 analyzer is a new automated lab-on-a-chip-based platelet function analyzer. OBJECTIVE: We aimed to evaluate a new platelet function analyzing system, the Anysis-200 in comparison to the Platelet Function Analyzer (PFA)-200 in cardiac patients. METHODS: Citrated blood was collected from 174 patients who visited the Department of Cardiology. The Anysis-200 consists of two kits, the microchips with collagen and epinephrine-coated membrane (C/EPI) or adenosine diphosphate-coated membrane (C/ADP). Platelet clogging in the Anysis-200 is measured by the blood migration distance obtained by a camera, which is compatible with the closure time in the PFA-200. We performed Anysis-200 and PFA-200 analyzers simultaneously and compared the results. RESULTS: The sensitivity and specificity of the Anysis-200 C/EPI kit in comparison to the PFA-200 C/EPI kit were 63.41% and 91.43%, respectively. Regarding the C/ADP kit, the sensitivity and specificity of the Anysis-200 were 58.97% and 74.29%, respectively. The agreement rate between the Anysis-200 and PFA-200 for C/EPI was 83.35% and 70.14% for C/ADP. CONCLUSIONS: The Anysis-200, which applies a novel method to detect platelet clogging, has shown moderate to fair agreement with the PFA-200. This test is potentially useful for screening cardiac patients with an abnormal platelet function.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherIOS PRESS-
dc.subjectCLINICAL UTILITY-
dc.subjectAGGREGATION-
dc.titlePerformance evaluation of the Anysis-200 platelet function analyzer in cardiac patients-
dc.typeArticle-
dc.contributor.affiliatedAuthorSeo, Hong Seog-
dc.identifier.doi10.3233/CH-190801-
dc.identifier.scopusid2-s2.0-85094604816-
dc.identifier.wosid000752431800003-
dc.identifier.bibliographicCitationCLINICAL HEMORHEOLOGY AND MICROCIRCULATION, v.80, no.1, pp.17 - 24-
dc.relation.isPartOfCLINICAL HEMORHEOLOGY AND MICROCIRCULATION-
dc.citation.titleCLINICAL HEMORHEOLOGY AND MICROCIRCULATION-
dc.citation.volume80-
dc.citation.number1-
dc.citation.startPage17-
dc.citation.endPage24-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaHematology-
dc.relation.journalResearchAreaCardiovascular System & Cardiology-
dc.relation.journalWebOfScienceCategoryHematology-
dc.relation.journalWebOfScienceCategoryPeripheral Vascular Disease-
dc.subject.keywordPlusCLINICAL UTILITY-
dc.subject.keywordPlusAGGREGATION-
dc.subject.keywordAuthorPlatelet function-
dc.subject.keywordAuthorAnysis-200 analyzer-
dc.subject.keywordAuthorPFA-200 analyzer-