Inter-fractional entrance dose monitoring as quality assurance using Gafchromic EBT3 film

Authors
Moon, Sun YoungJo, YunhuiSeo, JaehyeonShin, DonghoYoon, Myonggeun
Issue Date
Jul-2022
Publisher
WOLTERS KLUWER MEDKNOW PUBLICATIONS
Keywords
Gafchromic EBT3 film; intensity-modulated radiotherapy; inter-fractional dosimetry; three-dimensional conformal radiotherapy; treatment quality assurance
Citation
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, v.18, no.4, pp 1152 - 1158
Pages
7
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS
Volume
18
Number
4
Start Page
1152
End Page
1158
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/190375
DOI
10.4103/jcrt.JCRT_8_20
ISSN
0973-1482
1998-4138
Abstract
Introduction: This study describes a simple method of inter-fractional photon beam monitoring to measure the entrance dose of radiation treatment using Gafchromic EBT3 film. Materials and Methods: The film was placed at the center of a 1-cm thick phantom shaped like a block tray and fixed on the accessory tray of the gantry. The entrance dose was measured following the placement of the film in the accessory tray. The dose distribution calculated with the treatment planning system was compared with the dose distribution on the irradiated EBT3 films. The effectiveness of this methodology, as determined by gamma passing rates, was evaluated for the 22 fields of eight three-dimensional conformal radiotherapy (3D-CRT) plans and the 41 fields of nine intensity-modulated radiotherapy (RT) plans. The plans for three-dimensional conformal RT included treatments of the rectum, liver, breast, and head and neck, whereas the plans for intensity-modulated RT included treatments of the liver, brain, and lung. Results: The gamma passing rates for 3D-CRT ranged from 96.4% to 99.5%, with the mean gamma passing rate for 22 fields being 98.0%. The gamma passing rate for intensity-modulated RT ranged from 96.1% to 98.9%, with the mean gamma passing rate for 41 fields being 97.7%. All gamma indices were over the 95% tolerance level. Conclusions: The methodology described in this study, based on Gafchromic EBT3 film, can be utilized for inter-fractional entrance dose monitoring as quality assurance during RT. Clinical application of this method to patients can verify the accuracy of beam delivery in the treatment room.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Myong geun photo

Yoon, Myong geun
College of Health Sciences (School of Biomedical Engineering)
Read more

Altmetrics

Total Views & Downloads

BROWSE