USE OF CLINICAL EXPOSURE INDEX AND DEVIATION INDEX BASED ON NATIONAL DIAGNOSTIC REFERENCE LEVEL AS DOSE-OPTIMIZATION TOOLS FOR GENERAL RADIOGRAPHY IN KOREA
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Park, Hyemin | - |
dc.contributor.author | Yoon, Yongsu | - |
dc.contributor.author | Kim, Jungmin | - |
dc.contributor.author | Kim, Jungsu | - |
dc.contributor.author | Jeong, Hoiwoun | - |
dc.contributor.author | Tanaka, Nobukazu | - |
dc.contributor.author | Morishita, Junji | - |
dc.date.accessioned | 2021-08-30T13:04:11Z | - |
dc.date.available | 2021-08-30T13:04:11Z | - |
dc.date.created | 2021-06-18 | - |
dc.date.issued | 2020-10 | - |
dc.identifier.issn | 0144-8420 | - |
dc.identifier.uri | https://scholar.korea.ac.kr/handle/2021.sw.korea/52685 | - |
dc.description.abstract | The International Electrotechnical Commission introduced the concepts of exposure index (EI), target exposure index (EIT) and deviation index (DI) to manage and optimize patient dose in real time. In this study, we have proposed an appropriate method for setting the EIT based on the Korean national diagnostic reference levels (DRLs). Furthermore, we evaluated the use of clinical EI, EIT and DI as tools for patient dose optimization in clinical environments by observing the changes in DI with those in EIT. According to the Korean national exposure conditions, we conducted experiments on three representative radiographic examinations (chest posterior-anterior, lateral and abdomen anterior-posterior) of clinical environments. As the exposure conditions and DRLs varied, the clinical EI, EIT and DI also varied. These results reveal that the clinical EI, EIT and DI can be used as tools for optimizing the patient dose if EIT is periodically and properly updated. | - |
dc.language | English | - |
dc.language.iso | en | - |
dc.publisher | OXFORD UNIV PRESS | - |
dc.subject | DIGITAL RADIOGRAPHY | - |
dc.title | USE OF CLINICAL EXPOSURE INDEX AND DEVIATION INDEX BASED ON NATIONAL DIAGNOSTIC REFERENCE LEVEL AS DOSE-OPTIMIZATION TOOLS FOR GENERAL RADIOGRAPHY IN KOREA | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | Kim, Jungmin | - |
dc.identifier.doi | 10.1093/rpd/ncaa185 | - |
dc.identifier.scopusid | 2-s2.0-85099125939 | - |
dc.identifier.wosid | 000606028800006 | - |
dc.identifier.bibliographicCitation | RADIATION PROTECTION DOSIMETRY, v.191, no.4, pp.439 - 451 | - |
dc.relation.isPartOf | RADIATION PROTECTION DOSIMETRY | - |
dc.citation.title | RADIATION PROTECTION DOSIMETRY | - |
dc.citation.volume | 191 | - |
dc.citation.number | 4 | - |
dc.citation.startPage | 439 | - |
dc.citation.endPage | 451 | - |
dc.type.rims | ART | - |
dc.type.docType | Article | - |
dc.description.journalClass | 1 | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | Environmental Sciences & Ecology | - |
dc.relation.journalResearchArea | Public, Environmental & Occupational Health | - |
dc.relation.journalResearchArea | Nuclear Science & Technology | - |
dc.relation.journalResearchArea | Radiology, Nuclear Medicine & Medical Imaging | - |
dc.relation.journalWebOfScienceCategory | Environmental Sciences | - |
dc.relation.journalWebOfScienceCategory | Public, Environmental & Occupational Health | - |
dc.relation.journalWebOfScienceCategory | Nuclear Science & Technology | - |
dc.relation.journalWebOfScienceCategory | Radiology, Nuclear Medicine & Medical Imaging | - |
dc.subject.keywordPlus | DIGITAL RADIOGRAPHY | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.