Performance evaluation of three i-SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Kim, Ha Nui | - |
dc.contributor.author | Moon, Kyung Chul | - |
dc.contributor.author | Yoon, Soo-Young | - |
dc.date.accessioned | 2021-08-30T18:05:57Z | - |
dc.date.available | 2021-08-30T18:05:57Z | - |
dc.date.created | 2021-06-19 | - |
dc.date.issued | 2020-08 | - |
dc.identifier.issn | 0887-8013 | - |
dc.identifier.uri | https://scholar.korea.ac.kr/handle/2021.sw.korea/53884 | - |
dc.description.abstract | Background Capillary blood is the most commonly used sample for point-of-care (POC) glucometers. However, in critically ill patients, the glucose levels measured from capillary blood may not be reliable. Thus, we aimed to evaluate and compare the accuracy of glucose levels measured with POC glucometers and the YSI 2300 glucose analyzer using leftover arterial blood samples. Methods In total, 100 leftover heparinized arterial blood samples were used to evaluate the performance of three i-SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU-CHEK(R) Inform II glucometer. The reference value was obtained using the YSI 2300 glucose analyzer. The results were analyzed based on International Organization for Standardization 15197:2013 guidelines. Results More than 95% of results obtained using POC glucometers were within +/- 15 mg/dL of the reference value for glucose concentrations <100 mg/dL and within +/- 15% of the reference value for glucose concentrations >= 100 mg/dL. In the consensus error grid analysis, more than 99% of results were found to be within zones A and B. An excellent correlation was found between the values obtained using POC glucometers and the YSI 2300 glucose analyzer (R-2 > .99). Conclusion The i-SENS glucometers showed stable and accurate results when leftover arterial blood samples were used. Therefore, POC glucometers could be useful in critical care settings, such as intensive care units, where arterial samples are routinely used. | - |
dc.language | English | - |
dc.language.iso | en | - |
dc.publisher | WILEY | - |
dc.subject | METERS | - |
dc.title | Performance evaluation of three i-SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | Yoon, Soo-Young | - |
dc.identifier.doi | 10.1002/jcla.23356 | - |
dc.identifier.scopusid | 2-s2.0-85085484300 | - |
dc.identifier.wosid | 000534054300001 | - |
dc.identifier.bibliographicCitation | JOURNAL OF CLINICAL LABORATORY ANALYSIS, v.34, no.8 | - |
dc.relation.isPartOf | JOURNAL OF CLINICAL LABORATORY ANALYSIS | - |
dc.citation.title | JOURNAL OF CLINICAL LABORATORY ANALYSIS | - |
dc.citation.volume | 34 | - |
dc.citation.number | 8 | - |
dc.type.rims | ART | - |
dc.type.docType | Article | - |
dc.description.journalClass | 1 | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | Medical Laboratory Technology | - |
dc.relation.journalWebOfScienceCategory | Medical Laboratory Technology | - |
dc.subject.keywordPlus | METERS | - |
dc.subject.keywordAuthor | arterial blood i-SENS | - |
dc.subject.keywordAuthor | glucometer | - |
dc.subject.keywordAuthor | POCT | - |
dc.subject.keywordAuthor | point-of-care | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.