Influence of implant-abutment connection structure on peri-implant bone level in a second molar: A 1-year randomized controlled trial
- Authors
- Kim, Jin-Cheol; Lee, Jungwon; Kim, Sungtae; Koo, Ki-Tae; Kim, Hae-Young; Yeo, In-Sung Luke
- Issue Date
- 6월-2019
- Publisher
- KOREAN ACAD PROSTHODONTICS
- Keywords
- Randomized controlled trial; Dental implant-abutment design; Alveolar bone loss
- Citation
- JOURNAL OF ADVANCED PROSTHODONTICS, v.11, no.3, pp.147 - 154
- Indexed
- SCIE
SCOPUS
KCI
- Journal Title
- JOURNAL OF ADVANCED PROSTHODONTICS
- Volume
- 11
- Number
- 3
- Start Page
- 147
- End Page
- 154
- URI
- https://scholar.korea.ac.kr/handle/2021.sw.korea/65218
- DOI
- 10.4047/jap.2019.11.3.147
- ISSN
- 2005-7806
- Abstract
- PURPOSE. This study aimed to evaluate the effect of two different implant-abutment connection structures with identical implant design on peri-implant bone level. MATERIALS AND METHODS. This clinical study was a patient-blind randomized controlled trial following the CONSORT 2010 checklists. This trial was conducted in 24 patients recruited between March 2013 and July 2015. Implants with internal friction connection were compared to those with external hex connection. One implant for each patient was installed, replacing the second molar. Implant-supported crowns were delivered at four months after implant insertion. Standardized periapical radiographs were taken at prosthesis delivery (baseline), and one year after delivery. On the radiographs, distance from implant shoulder to first bone-to-implant contact (DIB) and peri-implant area were measured, which were the primary and secondary outcome, respectively. RESULTS. Eleven external and eleven internal implants were analyzed. Mean changes of DIB from baseline to 1-year postloading were 0.59 (0.95) mm for the external and 0.01 (0.68) mm for the internal connection. Although no significant differences were found between the two groups, medium effect size was found in DIB between the connections (Cohen's d = 0.67). CONCLUSION. Considering the effect size in DIB, this study suggested the possibility of the internal friction connection structure for more effective preservation of marginal bone.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - College of Health Sciences > Division of Health Policy and Management > 1. Journal Articles
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.