Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Effect of metallic tools on scattered radiation dose during the use of C-arm fluoroscopy in orthopaedic surgery

Full metadata record
DC Field Value Language
dc.contributor.authorSung, Ki Hyuk-
dc.contributor.authorJung, Young-Jun-
dc.contributor.authorCha, Hyemi-
dc.contributor.authorChung, Chin Youb-
dc.contributor.authorLee, Kisung-
dc.contributor.authorPark, Moon Seok-
dc.date.accessioned2021-09-01T21:56:25Z-
dc.date.available2021-09-01T21:56:25Z-
dc.date.created2021-06-19-
dc.date.issued2019-01-
dc.identifier.issn0449-3060-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/68461-
dc.description.abstractThis study investigated the effect of metallic tools on the scattered radiation dose delivered to surgeons' radiosensitive organs while simulating hip surgery using C-arm fluoroscopy. Two phantoms, a pelvis and a Rando phantom, were used to simulate a patient and a surgeon in this study. Photoluminescence dosimeters were inserted into the Rando phantom in the positions of the eye, thyroid and gonad. A drill was positioned above the hip of the pelvis phantom or beside the pelvis phantom of the same height. For each drill location, the scattered radiation dose was measured when the angle to the operator phantom was 45 degrees; this was repeated when the angle was 90 degrees. The scattered radiation doses to the eye, thyroid and gonad when the drill was placed beside the pelvis phantom with 90 degrees angulation to the operator phantom were significantly lower than the reference values and those when the drill was placed beside the pelvis phantom at a 45 degrees angulation to the operator phantom. The scattered radiation doses to the eye and thyroid when the drill was placed above the hip were significantly lower than the references values. Of the four different scenarios, the scattered radiation doses to the eye, thyroid and gonad were lowest when the drill was placed beside the pelvis phantom with 90 degrees angulation. This study showed that the scattered radiation doses to radiosensitive organs were affected by the location and angle of the metallic tools in relation to the operator. Therefore, orthopedic surgeons should consider the effect of metallic tools on the scattered radiation dose during intraoperative use of C-arm fluoroscopy.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherOXFORD UNIV PRESS-
dc.subjectIONIZING-RADIATION-
dc.subjectINTRAOPERATIVE USE-
dc.subjectEXPOSURE-
dc.subjectMINI-
dc.subjectBACKSCATTER-
dc.subjectINTERFACES-
dc.subjectSTANDARD-
dc.titleEffect of metallic tools on scattered radiation dose during the use of C-arm fluoroscopy in orthopaedic surgery-
dc.typeArticle-
dc.contributor.affiliatedAuthorLee, Kisung-
dc.identifier.doi10.1093/jrr/rry073-
dc.identifier.scopusid2-s2.0-85061590754-
dc.identifier.wosid000462583100001-
dc.identifier.bibliographicCitationJOURNAL OF RADIATION RESEARCH, v.60, no.1, pp.1 - 6-
dc.relation.isPartOfJOURNAL OF RADIATION RESEARCH-
dc.citation.titleJOURNAL OF RADIATION RESEARCH-
dc.citation.volume60-
dc.citation.number1-
dc.citation.startPage1-
dc.citation.endPage6-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaLife Sciences & Biomedicine - Other Topics-
dc.relation.journalResearchAreaOncology-
dc.relation.journalResearchAreaRadiology, Nuclear Medicine & Medical Imaging-
dc.relation.journalWebOfScienceCategoryBiology-
dc.relation.journalWebOfScienceCategoryOncology-
dc.relation.journalWebOfScienceCategoryRadiology, Nuclear Medicine & Medical Imaging-
dc.subject.keywordPlusIONIZING-RADIATION-
dc.subject.keywordPlusINTRAOPERATIVE USE-
dc.subject.keywordPlusEXPOSURE-
dc.subject.keywordPlusMINI-
dc.subject.keywordPlusBACKSCATTER-
dc.subject.keywordPlusINTERFACES-
dc.subject.keywordPlusSTANDARD-
dc.subject.keywordAuthorscattered radiation dose-
dc.subject.keywordAuthorradiosensitive organ-
dc.subject.keywordAuthorC-arm fluoroscopy-
dc.subject.keywordAuthormetallic tool-
Files in This Item
There are no files associated with this item.
Appears in
Collections
Graduate School > Department of Bioengineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Ki sung photo

Lee, Ki sung
바이오의공학과
Read more

Altmetrics

Total Views & Downloads

BROWSE