Image quality assessment with dose reduction using high kVp and additional filtration for abdominal digital radiography

Full metadata record
DC Field Value Language
dc.contributor.authorJang, Ji Sung-
dc.contributor.authorYang, Hyung Jin-
dc.contributor.authorKoo, Hyun Jung-
dc.contributor.authorKim, Sung Ho-
dc.contributor.authorPark, Chan Rok-
dc.contributor.authorYoon, Suk Hwan-
dc.contributor.authorShin, So Youn-
dc.contributor.authorDo, Kyung-Hyun-
dc.date.accessioned2021-09-02T10:40:14Z-
dc.date.available2021-09-02T10:40:14Z-
dc.date.created2021-06-19-
dc.date.issued2018-06-
dc.identifier.issn1120-1797-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/75062-
dc.description.abstractPurpose: Dose reduction using additional filters with high kilovoltage peak (kVp) for abdominal digital radiography has received much attention recently. We evaluated image quality with dose reduction in abdominal digital radiography by using high kVp and additional copper filters at a tertiary hospital. Methods: Between June 2016 and July 2016, 82 patients underwent abdominal digital radiography using 80 kVp in X-ray room 1 and 82 were imaged using 92 kVp with 0.1-mm copper filtration in X-ray room 2. The effective dose was calculated using a PC-based Monte Carlo program. Image quality of the abdominal radiography acquired in the two rooms was evaluated using a five-point ordinal scale, as well as the signal-to-noise and contrast-to-noise ratios. Results: The mean effective dose decreased by 25.8% and 25.7% for the supine and standing positions, respectively, when abdominal digital radiography using 92kVp with 0.1-mm copper filtration was performed. In the 20 patients who performed abdominal digital radiography twice in each room, visual grading scores for visualisation of psoas outlines and kidney outlines are higher in room 1. However, there was no statistical significant difference of visual grading scores among the 124 patients who underwent only one abdominal radiography in the room 1 or 2 (P > 0.05). Conclusions: Dose reduction for abdominal digital radiography can be achieved with comparable image quality by performing abdominal digital radiography using 92kVp with 0.1-mm copper filtration, despite the higher AEC dose.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherELSEVIER SCI LTD-
dc.subjectFLAT-PANEL DETECTOR-
dc.subjectTO-NOISE RATIO-
dc.subjectCHEST RADIOGRAPHY-
dc.subjectCOMPUTED RADIOGRAPHY-
dc.subjectCOPPER FILTRATION-
dc.subjectTUBE VOLTAGE-
dc.subjectRADIOLOGY-
dc.subjectRADIATION-
dc.titleImage quality assessment with dose reduction using high kVp and additional filtration for abdominal digital radiography-
dc.typeArticle-
dc.contributor.affiliatedAuthorYang, Hyung Jin-
dc.identifier.doi10.1016/j.ejmp.2018.05.007-
dc.identifier.scopusid2-s2.0-85047442753-
dc.identifier.wosid000436777000007-
dc.identifier.bibliographicCitationPHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, v.50, pp.46 - 51-
dc.relation.isPartOfPHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS-
dc.citation.titlePHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS-
dc.citation.volume50-
dc.citation.startPage46-
dc.citation.endPage51-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaRadiology, Nuclear Medicine & Medical Imaging-
dc.relation.journalWebOfScienceCategoryRadiology, Nuclear Medicine & Medical Imaging-
dc.subject.keywordPlusFLAT-PANEL DETECTOR-
dc.subject.keywordPlusTO-NOISE RATIO-
dc.subject.keywordPlusCHEST RADIOGRAPHY-
dc.subject.keywordPlusCOMPUTED RADIOGRAPHY-
dc.subject.keywordPlusCOPPER FILTRATION-
dc.subject.keywordPlusTUBE VOLTAGE-
dc.subject.keywordPlusRADIOLOGY-
dc.subject.keywordPlusRADIATION-
dc.subject.keywordAuthorAbdominal radiography-
dc.subject.keywordAuthorEffective dose-
dc.subject.keywordAuthorDose reduction-
dc.subject.keywordAuthorCopper filter-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Science and Technology > Semiconductor Physics in Division of Display and Semiconductor Physics > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE