Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Statistical process control analysis for patient quality assurance of intensity modulated radiation therapy

Full metadata record
DC Field Value Language
dc.contributor.authorLee, Rena-
dc.contributor.authorKim, Kyubo-
dc.contributor.authorCho, Samju-
dc.contributor.authorLim, Sangwook-
dc.contributor.authorLee, Suk-
dc.contributor.authorShim, Jang Bo-
dc.contributor.authorHuh, Hyun Do-
dc.contributor.authorLee, Sang Hoon-
dc.contributor.authorAhn, Sohyun-
dc.date.accessioned2021-09-02T23:26:13Z-
dc.date.available2021-09-02T23:26:13Z-
dc.date.created2021-06-19-
dc.date.issued2017-11-
dc.identifier.issn0374-4884-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/81712-
dc.description.abstractThis study applied statistical process control to set and verify the quality assurances (QA) tolerance standard for our hospital's characteristics with the criteria standards that are applied to all the treatment sites with this analysis. Gamma test factor of delivery quality assurances (DQA) was based on 3%/3 mm. Head and neck, breast, prostate cases of intensity modulated radiation therapy (IMRT) or volumetric arc radiation therapy (VMAT) were selected for the analysis of the QA treatment sites. The numbers of data used in the analysis were 73 and 68 for head and neck patients. Prostate and breast were 49 and 152 by MapCHECK and ArcCHECK respectively. C (p) value of head and neck and prostate QA were above 1.0, C (pml) is 1.53 and 1.71 respectively, which is close to the target value of 100%. C (pml) value of breast (IMRT) was 1.67, data values are close to the target value of 95%. But value of was 0.90, which means that the data values are widely distributed. C (p) and C (pml) of breast VMAT QA were respectively 1.07 and 2.10. This suggests that the VMAT QA has better process capability than the IMRT QA. Consequently, we should pay more attention to planning and QA before treatment for breast Radiotherapy.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherKOREAN PHYSICAL SOC-
dc.titleStatistical process control analysis for patient quality assurance of intensity modulated radiation therapy-
dc.typeArticle-
dc.contributor.affiliatedAuthorLee, Suk-
dc.identifier.doi10.3938/jkps.71.717-
dc.identifier.scopusid2-s2.0-85035084709-
dc.identifier.wosid000415755100017-
dc.identifier.bibliographicCitationJOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.71, no.10, pp.717 - 721-
dc.relation.isPartOfJOURNAL OF THE KOREAN PHYSICAL SOCIETY-
dc.citation.titleJOURNAL OF THE KOREAN PHYSICAL SOCIETY-
dc.citation.volume71-
dc.citation.number10-
dc.citation.startPage717-
dc.citation.endPage721-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.identifier.kciidART002283748-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.description.journalRegisteredClasskci-
dc.relation.journalResearchAreaPhysics-
dc.relation.journalWebOfScienceCategoryPhysics, Multidisciplinary-
dc.subject.keywordAuthorStatistical process control-
dc.subject.keywordAuthorProcess capability index-
dc.subject.keywordAuthorGamma index-
dc.subject.keywordAuthorGamma pass rate-
dc.subject.keywordAuthorIMRT/VMAT QA-