dc.description.abstract | Purpose: The purpose of this study was to evaluate the internal fitness of the resin coping that was fabricated by the traditional and Digital manufacturing methods through 3-dimensional analysis. Methods: maxillary right second molar was chosen implant master model. Custom-built impression trays were manufactured. After screwing the pick-up impression coping onto the master cast, impressions were made with silicone impression. The Working model was then made with type IV stone. The coping was fabricated: SLAC group (n=8), APPC group (n=8), LAPC group (n=8) Resin coping data was measured by using a three-dimensional evaluation program. Internal fitness was calculated by RMS (Root Mean Square).It measures mean and Stand Deviation (SD). Results: Three groups are measured 47.11±(3.08)㎛ total RMS of SLAC group, 48.35(±1.55㎛) for total RMS of LAPC group, 43.45±2.09㎛ for total RMS of APPC group. Measured value is gradually increased. Followed by autopolymerized pattern resin; Stereolithography resin, Light-activated pattern resin But there were no differences stastically(P>0.321). Conclusion: Evaluation of internal fitness on Resin copings was fabricated by three–ways methods showed that no differences statistically significant and clinically acceptable results | - |
dc.language | Korean | - |
dc.language.iso | ko | - |
dc.publisher | 대한치과기공학회 | - |
dc.title | 세 가지 방식으로 제작된 레진코핑의 내면적합성 평가: 3차원적 분석 | - |
dc.title.alternative | Assessment of internal fitness on resin copings fabricated by 3 – ways methods: Three - dimensional analysis | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | 김웅철 | - |
dc.identifier.doi | 10.14347/kadt.2017.39.1.17 | - |
dc.identifier.bibliographicCitation | 대한치과기공학회지, v.39, no.1, pp.17 - 24 | - |
dc.relation.isPartOf | 대한치과기공학회지 | - |
dc.citation.title | 대한치과기공학회지 | - |
dc.citation.volume | 39 | - |
dc.citation.number | 1 | - |
dc.citation.startPage | 17 | - |
dc.citation.endPage | 24 | - |
dc.type.rims | ART | - |
dc.identifier.kciid | ART002211265 | - |
dc.description.journalClass | 2 | - |
dc.description.journalRegisteredClass | kci | - |
dc.description.journalRegisteredClass | other | - |
dc.subject.keywordAuthor | Denture | - |
dc.subject.keywordAuthor | Glass fiber | - |
dc.subject.keywordAuthor | Resin | - |
dc.subject.keywordAuthor | Shrinkage | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.