Monitoring 3D dose distributions in proton therapy by reconstruction using an iterative method
- Authors
- Kim, Young-Hak; Yoon, Changyeon; Lee, Wonho
- Issue Date
- 8월-2016
- Publisher
- PERGAMON-ELSEVIER SCIENCE LTD
- Keywords
- Proton therapy; Bragg peak; Prompt gamma-Rays; Single-Photon Emission Computed Tomography (SPECT); Iterative reconstruction method
- Citation
- APPLIED RADIATION AND ISOTOPES, v.114, pp.33 - 39
- Indexed
- SCIE
SCOPUS
- Journal Title
- APPLIED RADIATION AND ISOTOPES
- Volume
- 114
- Start Page
- 33
- End Page
- 39
- URI
- https://scholar.korea.ac.kr/handle/2021.sw.korea/87978
- DOI
- 10.1016/j.apradiso.2016.05.005
- ISSN
- 0969-8043
- Abstract
- The Bragg peak of protons can be determined by measuring prompt gamma-rays. In this study, prompt gamma-rays detected by single-photon emission computed tomography with a geometrically optimized collimation system were reconstructed by an iterative method. The falloff position by iterative method (52.48 mm) was most similar to the Bragg peak (52 mm) of an 80 MeV proton compared with those of back-projection (54.11 mm) and filtered back-projection (54.91 mm) methods. Iterative method also showed better image performance than other methods. (C) 2016 Elsevier Ltd. All rights reserved.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - College of Health Sciences > School of Health and Environmental Science > 1. Journal Articles
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.