Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Feasibility Study of Patient-specific Quality Assurance System for High-dose-rate Brachytherapy in Patients with Cervical Cancer

Full metadata record
DC Field Value Language
dc.contributor.authorLee, Boram-
dc.contributor.authorAhn, Sung Hwan-
dc.contributor.authorKim, Hyeyoung-
dc.contributor.authorHan, Youngyih-
dc.contributor.authorHuh, Seung Jae-
dc.contributor.authorKim, Jin Sung-
dc.contributor.authorKim, Dong Wook-
dc.contributor.authorSim, Jina-
dc.contributor.authorYoon, Myonggeun-
dc.date.accessioned2021-09-04T01:13:04Z-
dc.date.available2021-09-04T01:13:04Z-
dc.date.created2021-06-17-
dc.date.issued2016-04-
dc.identifier.issn0374-4884-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/89128-
dc.description.abstractThis study was conducted for the purpose of establishing a quality-assurance (QA) system for brachytherapy that can ensure patient-specific QA by enhancing dosimetric accuracy for the patient's therapy plan. To measure the point-absorbed dose and the 2D dose distribution for the patient's therapy plan, we fabricated a solid phantom that allowed for the insertion of an applicator for patient-specific QA and used an ion chamber and a film as measuring devices. The patient treatment plan was exported to the QA dose-calculation software, which calculated the time weight of dwell position stored in the plan DICOM (Digital Imaging and Communications in Medicine) file to obtain an overall beam quality correction factor, and that correction was applied to the dose calculations. Experiments were conducted after importing the patient's treatment planning source data for the fabricated phantom and inserting the applicator, ion chamber, and film into the phantom. On completion of dose delivery, the doses to the ion chamber and film were checked against the corresponding treatment plan to evaluate the dosimetric accuracy. For experimental purposes, five treatment plans were randomly selected. The beam quality correction factors for ovoid and tandem brachytherapy applicators were found to be 1.15 and 1.10 - 1.12, respectively. The beam quality correction factor in tandem fluctuated by approximately 2%, depending on the changes in the dwell position. The doses measured by using the ion chamber showed differences ranging from -2.4% to 0.6%, compared to the planned doses. As for the film, the passing rate was 90% or higher when assessed using a gamma value of the local dose difference of 3% and a distance to agreement of 3 mm. The results show that the self-fabricated phantom was suitable for QA in clinical settings. The proposed patient-specific QA for the treatment planning is expected to contribute to reduce dosimetric errors in brachytherapy and, thus, to enhancing treatment accuracy.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherKOREAN PHYSICAL SOC-
dc.subjectIMAGE-GUIDED BRACHYTHERAPY-
dc.subjectWORKING GROUP-
dc.subjectUTERINE PERFORATION-
dc.subjectINTRACAVITARY BRACHYTHERAPY-
dc.subjectAPPLICATOR RECONSTRUCTION-
dc.subjectCOMPUTED-TOMOGRAPHY-
dc.subjectCARCINOMA-
dc.subjectRECOMMENDATIONS-
dc.subjectDOSIMETRY-
dc.subjectPHANTOM-
dc.titleFeasibility Study of Patient-specific Quality Assurance System for High-dose-rate Brachytherapy in Patients with Cervical Cancer-
dc.typeArticle-
dc.contributor.affiliatedAuthorYoon, Myonggeun-
dc.identifier.doi10.3938/jkps.68.1029-
dc.identifier.scopusid2-s2.0-84964311344-
dc.identifier.wosid000374334500017-
dc.identifier.bibliographicCitationJOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.68, no.8, pp.L1029 - L1036-
dc.relation.isPartOfJOURNAL OF THE KOREAN PHYSICAL SOCIETY-
dc.citation.titleJOURNAL OF THE KOREAN PHYSICAL SOCIETY-
dc.citation.volume68-
dc.citation.number8-
dc.citation.startPageL1029-
dc.citation.endPageL1036-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.identifier.kciidART002101182-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.description.journalRegisteredClasskci-
dc.relation.journalResearchAreaPhysics-
dc.relation.journalWebOfScienceCategoryPhysics, Multidisciplinary-
dc.subject.keywordPlusIMAGE-GUIDED BRACHYTHERAPY-
dc.subject.keywordPlusWORKING GROUP-
dc.subject.keywordPlusUTERINE PERFORATION-
dc.subject.keywordPlusINTRACAVITARY BRACHYTHERAPY-
dc.subject.keywordPlusAPPLICATOR RECONSTRUCTION-
dc.subject.keywordPlusCOMPUTED-TOMOGRAPHY-
dc.subject.keywordPlusCARCINOMA-
dc.subject.keywordPlusRECOMMENDATIONS-
dc.subject.keywordPlusDOSIMETRY-
dc.subject.keywordPlusPHANTOM-
dc.subject.keywordAuthorBrachytherapy-
dc.subject.keywordAuthorPatient-specific QA-
dc.subject.keywordAuthorApplicator-
dc.subject.keywordAuthorPhantom-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Myong geun photo

Yoon, Myong geun
보건과학대학 (바이오의공학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE