Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

What is the underlying mechanism for the failure mode observed in the proximal femoral locking compression plate? A biomechanical study

Full metadata record
DC Field Value Language
dc.contributor.authorSchneider, Kerstin-
dc.contributor.authorOh, Jong-Keon-
dc.contributor.authorZderic, Ivan-
dc.contributor.authorStoffel, Karl-
dc.contributor.authorRichards, R. Geoff-
dc.contributor.authorWolf, Stefan-
dc.contributor.authorGueorguiev, Boyko-
dc.contributor.authorNork, Sean E.-
dc.date.accessioned2021-09-04T13:47:05Z-
dc.date.available2021-09-04T13:47:05Z-
dc.date.created2021-06-18-
dc.date.issued2015-08-
dc.identifier.issn0020-1383-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/92834-
dc.description.abstractPurpose: Recently, several cases of clinical failure have been reported for the Proximal Femoral Locking Compression Plate (PF-LCP). The current study was designed to explore biomechanically the underlying mechanism and to determine whether the observed failure was due to technical error on insertion or to implant design. Methods: A foam block model simulating an unstable intertrochanteric fracture was created for 3 study groups with 6 specimens each. Group C was correctly instrumented according to the manufacturer's guidelines. In Group P and Group A, the first or second proximal screw was placed with a posterior or anterior off-axis orientation by 28 measured in the transversal plane, respectively. Each construct was cyclically tested until failure using a test setup and protocol simulating complex axial and torsional loading. Radiographs were taken prior to and after the tests. Force, number of cycles to failure and failure mode were compared. Results: A screw deviation of 28 from the nominal axis led to significantly earlier construct failure in Group P and Group A in comparison to Group C. The failure mode was characterised by loosening of the off-axis screw due to disengagement with the plate, resulting in loss of construct stiffness and varus collapse of the fracture. Conclusions: In our biomechanical test setup, the clinical failure modes observed with the PF-LCP were reproducible. A screw deviation of 28 from the nominal axis consistently led to the failure. This highlights how crucial is the accurate placement of locking screws in the proximal femur. (C) 2015 Elsevier Ltd. All rights reserved.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherELSEVIER SCI LTD-
dc.subjectANGLED BLADE PLATE-
dc.subjectFEMUR FRACTURES-
dc.subjectCONTACT FORCES-
dc.subjectGAP MODEL-
dc.subjectFIXATION-
dc.subjectSCREW-
dc.subjectINSERTION-
dc.subjectNAIL-
dc.subjectSTABILITY-
dc.subjectGAIT-
dc.titleWhat is the underlying mechanism for the failure mode observed in the proximal femoral locking compression plate? A biomechanical study-
dc.typeArticle-
dc.contributor.affiliatedAuthorOh, Jong-Keon-
dc.identifier.doi10.1016/j.injury.2015.05.034-
dc.identifier.scopusid2-s2.0-84937526678-
dc.identifier.wosid000361894200010-
dc.identifier.bibliographicCitationINJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, v.46, no.8, pp.1483 - 1490-
dc.relation.isPartOfINJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED-
dc.citation.titleINJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED-
dc.citation.volume46-
dc.citation.number8-
dc.citation.startPage1483-
dc.citation.endPage1490-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaGeneral & Internal Medicine-
dc.relation.journalResearchAreaEmergency Medicine-
dc.relation.journalResearchAreaOrthopedics-
dc.relation.journalResearchAreaSurgery-
dc.relation.journalWebOfScienceCategoryCritical Care Medicine-
dc.relation.journalWebOfScienceCategoryEmergency Medicine-
dc.relation.journalWebOfScienceCategoryOrthopedics-
dc.relation.journalWebOfScienceCategorySurgery-
dc.subject.keywordPlusANGLED BLADE PLATE-
dc.subject.keywordPlusFEMUR FRACTURES-
dc.subject.keywordPlusCONTACT FORCES-
dc.subject.keywordPlusGAP MODEL-
dc.subject.keywordPlusFIXATION-
dc.subject.keywordPlusSCREW-
dc.subject.keywordPlusINSERTION-
dc.subject.keywordPlusNAIL-
dc.subject.keywordPlusSTABILITY-
dc.subject.keywordPlusGAIT-
dc.subject.keywordAuthorProximal femoral locking compression plate-
dc.subject.keywordAuthorUnstable trochanteric fracture-
dc.subject.keywordAuthorAngular stable plate fixation-
dc.subject.keywordAuthorFailure mode-
dc.subject.keywordAuthorBiomechanical study-
dc.subject.keywordAuthorScrew loosening-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Oh, Jong Keon photo

Oh, Jong Keon
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE