Electrical delay line multiplexing for pulsed mode radiation detectors
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Vinke, Ruud | - |
dc.contributor.author | Yeom, Jung Yeol | - |
dc.contributor.author | Levin, Craig S. | - |
dc.date.accessioned | 2021-09-04T17:22:12Z | - |
dc.date.available | 2021-09-04T17:22:12Z | - |
dc.date.created | 2021-06-18 | - |
dc.date.issued | 2015-04-07 | - |
dc.identifier.issn | 0031-9155 | - |
dc.identifier.uri | https://scholar.korea.ac.kr/handle/2021.sw.korea/93857 | - |
dc.description.abstract | Medical imaging systems are composed of a large number of position sensitive radiation detectors to provide high resolution imaging. For example, whole-body Positron Emission Tomography (PET) systems are typically composed of thousands of scintillation crystal elements, which are coupled to photosensors. Thus, PET systems greatly benefit from methods to reduce the number of data acquisition channels, in order to reduce the system development cost and complexity. In this paper we present an electrical delay line multiplexing scheme that can significantly reduce the number of readout channels, while preserving the signal integrity required for good time resolution performance. We experimented with two 4 x 4 LYSO crystal arrays, with crystal elements having 3 mm x 3 mm x 5 mm and 3 mm x 3 mm x 20 mm dimensions, coupled to 16 Hamamatsu MPPC S10931-050P SiPM elements. Results show that each crystal could be accurately identified, even in the presence of scintillation light sharing and inter-crystal Compton scatter among neighboring crystal elements. The multiplexing configuration degraded the coincidence timing resolution from similar to 243 ps FWHM to similar to 272 ps FWHM when 16 SiPM signals were combined into a single channel for the 4 x 4 LYSO crystal array with 3 mm x 3 mm x 20 mm crystal element dimensions, in coincidence with a 3 mm x 3 mm x 5 mm LYSO crystal pixel. The method is flexible to allow multiplexing configurations across different block detectors, and is scalable to an entire ring of detectors. | - |
dc.language | English | - |
dc.language.iso | en | - |
dc.publisher | IOP PUBLISHING LTD | - |
dc.subject | SILICON PHOTOMULTIPLIER | - |
dc.subject | AVALANCHE PHOTODIODES | - |
dc.subject | READOUT ELECTRONICS | - |
dc.subject | TIMING RESOLUTION | - |
dc.subject | TIME | - |
dc.title | Electrical delay line multiplexing for pulsed mode radiation detectors | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | Yeom, Jung Yeol | - |
dc.identifier.doi | 10.1088/0031-9155/60/7/2785 | - |
dc.identifier.scopusid | 2-s2.0-84925673059 | - |
dc.identifier.wosid | 000352516000013 | - |
dc.identifier.bibliographicCitation | PHYSICS IN MEDICINE AND BIOLOGY, v.60, no.7, pp.2785 - 2802 | - |
dc.relation.isPartOf | PHYSICS IN MEDICINE AND BIOLOGY | - |
dc.citation.title | PHYSICS IN MEDICINE AND BIOLOGY | - |
dc.citation.volume | 60 | - |
dc.citation.number | 7 | - |
dc.citation.startPage | 2785 | - |
dc.citation.endPage | 2802 | - |
dc.type.rims | ART | - |
dc.type.docType | Article | - |
dc.description.journalClass | 1 | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.relation.journalResearchArea | Engineering | - |
dc.relation.journalResearchArea | Radiology, Nuclear Medicine & Medical Imaging | - |
dc.relation.journalWebOfScienceCategory | Engineering, Biomedical | - |
dc.relation.journalWebOfScienceCategory | Radiology, Nuclear Medicine & Medical Imaging | - |
dc.subject.keywordPlus | SILICON PHOTOMULTIPLIER | - |
dc.subject.keywordPlus | AVALANCHE PHOTODIODES | - |
dc.subject.keywordPlus | READOUT ELECTRONICS | - |
dc.subject.keywordPlus | TIMING RESOLUTION | - |
dc.subject.keywordPlus | TIME | - |
dc.subject.keywordAuthor | radiation detectors | - |
dc.subject.keywordAuthor | multiplexing | - |
dc.subject.keywordAuthor | time-of-flight (TOF) | - |
dc.subject.keywordAuthor | positron emission tomography (PET) | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.