Feasibility Study of a Simple Approximation Algorithm for In-vivo Dose Reconstruction by Using the Transit Dose Measured Using an EPID
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Hwang, Ui-Jung | - |
dc.contributor.author | Song, Mi Hee | - |
dc.contributor.author | Baek, Tae Seong | - |
dc.contributor.author | Chung, Eun Ji | - |
dc.contributor.author | Yoon, Myonggeun | - |
dc.date.accessioned | 2021-09-04T19:44:39Z | - |
dc.date.available | 2021-09-04T19:44:39Z | - |
dc.date.created | 2021-06-15 | - |
dc.date.issued | 2015-02 | - |
dc.identifier.issn | 0374-4884 | - |
dc.identifier.uri | https://scholar.korea.ac.kr/handle/2021.sw.korea/94565 | - |
dc.description.abstract | The purpose of this study is to verify the accuracy of the dose delivered to the patient during intensity-modulated radiation therapy (IMRT) by using in-vivo dosimetry and to avoid accidental exposure to healthy tissues and organs close to tumors. The in-vivo dose was reconstructed by back projection of the transit dose with a simple approximation that considered only the percent depth dose and inverse square law. While the average gamma index for comparisons of dose distributions between the calculated dose map and the film measurement was less than the one for 96.3% of all pixels with the homogeneous phantom, the passing rate was reduced to 92.8% with the inhomogeneous phantom, suggesting that the reduction was apparently due to the inaccuracy of the reconstruction algorithm for inhomogeneity. The proposed method of calculating the dose inside a phantom was of comparable or better accuracy than the treatment planning system, suggesting that it can be used to verify the accuracy of the dose delivered to the patient, during treatment. | - |
dc.language | English | - |
dc.language.iso | en | - |
dc.publisher | KOREAN PHYSICAL SOC | - |
dc.subject | MODULATED RADIATION-THERAPY | - |
dc.subject | QUALITY-ASSURANCE | - |
dc.subject | VERIFICATION | - |
dc.subject | DOSIMETRY | - |
dc.subject | RADIOTHERAPY | - |
dc.title | Feasibility Study of a Simple Approximation Algorithm for In-vivo Dose Reconstruction by Using the Transit Dose Measured Using an EPID | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | Yoon, Myonggeun | - |
dc.identifier.doi | 10.3938/jkps.66.694 | - |
dc.identifier.scopusid | 2-s2.0-84924112992 | - |
dc.identifier.wosid | 000350893600026 | - |
dc.identifier.bibliographicCitation | JOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.66, no.4, pp.694 - 699 | - |
dc.relation.isPartOf | JOURNAL OF THE KOREAN PHYSICAL SOCIETY | - |
dc.citation.title | JOURNAL OF THE KOREAN PHYSICAL SOCIETY | - |
dc.citation.volume | 66 | - |
dc.citation.number | 4 | - |
dc.citation.startPage | 694 | - |
dc.citation.endPage | 699 | - |
dc.type.rims | ART | - |
dc.type.docType | Article | - |
dc.identifier.kciid | ART001965691 | - |
dc.description.journalClass | 1 | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.description.journalRegisteredClass | kci | - |
dc.relation.journalResearchArea | Physics | - |
dc.relation.journalWebOfScienceCategory | Physics, Multidisciplinary | - |
dc.subject.keywordPlus | MODULATED RADIATION-THERAPY | - |
dc.subject.keywordPlus | QUALITY-ASSURANCE | - |
dc.subject.keywordPlus | VERIFICATION | - |
dc.subject.keywordPlus | DOSIMETRY | - |
dc.subject.keywordPlus | RADIOTHERAPY | - |
dc.subject.keywordAuthor | Transit dose | - |
dc.subject.keywordAuthor | Back projection | - |
dc.subject.keywordAuthor | EPID | - |
dc.subject.keywordAuthor | Intensity modulated radiation therapy | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.