구강 내 스캐너와 구강 외 스캐너를 사용하여 취득된 스캔 데이터 정확도 비교Comparison of accuracy of digital data obtained by intraoral scanner and extra oral scanner

Other Titles
Comparison of accuracy of digital data obtained by intraoral scanner and extra oral scanner
Authors
이재준정일도김총명박진영김지환김웅철
Issue Date
2015
Publisher
대한치과기공학회
Keywords
Intra-oral scanner; Extra-oral scanner; Accuracy
Citation
대한치과기공학회지, v.37, no.4, pp.191 - 197
Indexed
KCI
OTHER
Journal Title
대한치과기공학회지
Volume
37
Number
4
Start Page
191
End Page
197
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/95266
DOI
10.14347/kadt.2015.37.4.191
ISSN
1229-3954
Abstract
Purpose: The purpose of this study was to compare the accuracy of the scan data acquired by the extra-oral and intra-oral scanner. Methods: The maxillary right first molar was made of polymethyl methacrylate (PMMA)specimen. This PMMA specimen was scanned with a engineering scanner and intra-oral scanner. Meanwhile, extra-oral scanner scanned stone die duplicated from PMMA master die. Trueness and precision of scan datas was measuredby3-dimensinal inspection. Independent t-test was conduct to analysis the significant difference.(a=0.05) Results: In the trueness analysis, mean of discrepancies were13.82um for intra oral scanner and16.84 um for extra-oral scanner. In the precision analysis, mean of discrepancies were 11.72 for inta-oral scanner and 9.2 for extra-oral scanner. Both trueness and precision showed a statistically significant difference (Table 1, p<0.05). Conclusion: Intra-oral scanner can show higher trueness than extra-oral scanner, it has lower precision.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > Division of Health Policy and Management > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE