치과 캐드캠 밀링장비에 따른 3본브릿지의 정확도 비교
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 배소연 | - |
dc.contributor.author | 박진영 | - |
dc.contributor.author | 김지환 | - |
dc.contributor.author | 김혜영 | - |
dc.contributor.author | 김명배 | - |
dc.contributor.author | 김웅철 | - |
dc.date.accessioned | 2021-09-04T23:46:45Z | - |
dc.date.available | 2021-09-04T23:46:45Z | - |
dc.date.created | 2021-06-17 | - |
dc.date.issued | 2015 | - |
dc.identifier.issn | 1229-3954 | - |
dc.identifier.uri | https://scholar.korea.ac.kr/handle/2021.sw.korea/95845 | - |
dc.description.abstract | Purpose: The purpose of this study was to compare the accuracy of the maxillary three-unit fixed dental prosthesis (FDPs) made using two CAD/CAM milling machines : DCM Group(Dentaim CAD/CAM milling machine), WCM Group(Wieland CAD/CAM milling machine). Methods: Each of 10 duplicate models was scanned by blue light scanner(Identica, Medit, Korea), and the threeunit FDPs (STL file) was designed using DelcamCAD. A total of 20 three-unit FDPs was fabricated, comprising 2 groups of 10 specimens each(shrinkage ratio is 1:1). The first three-unit FDPs STL file was used as a CAD reference model (CRM). Obtained STL files by scanning the inner surface of three-unit FDPs were converted into the point clouds-ASCⅡ files. Discrepancies between the point clouds and CRM were measured by superimposition software. Statistical methods to analyze the data were used non-parametric method. The mean (SD) values were compared by a Mann-Whitney U-test. Type one error rate was set at 0.05. Results: WCM group had small discrepancies with 2.17 of mean value compared to 4.44 in DCM group. The accuracy values between the two groups showed a statistically significant difference (Table 2, p<.05). Conclusion: The accuracy of the three-unit fixed dental prosthesis(FDPs) made of two CAD/CAM milling machines were statistically different. Accuracy with which the prosthesis made of WCM group was superior. | - |
dc.language | Korean | - |
dc.language.iso | ko | - |
dc.publisher | 대한치과기공학회 | - |
dc.title | 치과 캐드캠 밀링장비에 따른 3본브릿지의 정확도 비교 | - |
dc.title.alternative | The comparison of accuracy on three-unit fixed dental prosthesis made with CAD/CAM milling machines | - |
dc.type | Article | - |
dc.contributor.affiliatedAuthor | 김웅철 | - |
dc.identifier.doi | 10.14347/kadt.2014.37.1.9 | - |
dc.identifier.bibliographicCitation | 대한치과기공학회지, v.37, no.1, pp.9 - 15 | - |
dc.relation.isPartOf | 대한치과기공학회지 | - |
dc.citation.title | 대한치과기공학회지 | - |
dc.citation.volume | 37 | - |
dc.citation.number | 1 | - |
dc.citation.startPage | 9 | - |
dc.citation.endPage | 15 | - |
dc.type.rims | ART | - |
dc.identifier.kciid | ART002014124 | - |
dc.description.journalClass | 2 | - |
dc.description.journalRegisteredClass | kci | - |
dc.description.journalRegisteredClass | other | - |
dc.subject.keywordAuthor | Dental CAD/CAM system | - |
dc.subject.keywordAuthor | Accuracy | - |
dc.subject.keywordAuthor | Three-dimensional superimposition analysis | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
(02841) 서울특별시 성북구 안암로 14502-3290-1114
COPYRIGHT © 2021 Korea University. All Rights Reserved.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.