Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

Full metadata record
DC Field Value Language
dc.contributor.authorBaek, Tae Seong-
dc.contributor.authorChung, Eun Ji-
dc.contributor.authorSon, Jaeman-
dc.contributor.authorYoon, Myonggeun-
dc.date.accessioned2021-09-05T02:15:20Z-
dc.date.available2021-09-05T02:15:20Z-
dc.date.created2021-06-15-
dc.date.issued2014-12-04-
dc.identifier.issn1748-717X-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/96534-
dc.description.abstractPurpose: The aim of this study is to evaluate the ability of transit dosimetry using commercial treatment planning system (TPS) and an electronic portal imaging device (EPID) with simple calibration method to verify the beam delivery based on detection of large errors in treatment room. Methods and materials: Twenty four fields of intensity modulated radiotherapy (IMRT) plans were selected from four lung cancer patients and used in the irradiation of an anthropomorphic phantom. The proposed method was evaluated by comparing the calculated dose map from TPS and EPID measurement on the same plane using a gamma index method with a 3% dose and 3 mm distance-to-dose agreement tolerance limit. Results: In a simulation using a homogeneous plastic water phantom, performed to verify the effectiveness of the proposed method, the average passing rate of the transit dose based on gamma index was high enough, averaging 94.2% when there was no error during beam delivery. The passing rate of the transit dose for 24 IMRT fields was lower with the anthropomorphic phantom, averaging 86.8% +/- 3.8%, a reduction partially due to the inaccuracy of TPS calculations for inhomogeneity. Compared with the TPS, the absolute value of the transit dose at the beam center differed by -0.38% +/- 2.1%. The simulation study indicated that the passing rate of the gamma index was significantly reduced, to less than 40%, when a wrong field was erroneously irradiated to patient in the treatment room. Conclusions: This feasibility study suggested that transit dosimetry based on the calculation with commercial TPS and EPID measurement with simple calibration can provide information about large errors for treatment beam delivery.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherBMC-
dc.subjectPRETREATMENT VERIFICATION-
dc.subjectRADIOTHERAPY-
dc.subjectOPTIMIZATION-
dc.subjectWEIGHTS-
dc.titleFeasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry-
dc.typeArticle-
dc.contributor.affiliatedAuthorYoon, Myonggeun-
dc.identifier.doi10.1186/s13014-014-0273-8-
dc.identifier.scopusid2-s2.0-84924093272-
dc.identifier.wosid000349262000001-
dc.identifier.bibliographicCitationRADIATION ONCOLOGY, v.9-
dc.relation.isPartOfRADIATION ONCOLOGY-
dc.citation.titleRADIATION ONCOLOGY-
dc.citation.volume9-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaOncology-
dc.relation.journalResearchAreaRadiology, Nuclear Medicine & Medical Imaging-
dc.relation.journalWebOfScienceCategoryOncology-
dc.relation.journalWebOfScienceCategoryRadiology, Nuclear Medicine & Medical Imaging-
dc.subject.keywordPlusPRETREATMENT VERIFICATION-
dc.subject.keywordPlusRADIOTHERAPY-
dc.subject.keywordPlusOPTIMIZATION-
dc.subject.keywordPlusWEIGHTS-
dc.subject.keywordAuthorTransit dose-
dc.subject.keywordAuthorEPID-
dc.subject.keywordAuthorIntensity modulated radiation therapy-
dc.subject.keywordAuthorGamma index-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Myong geun photo

Yoon, Myong geun
보건과학대학 (바이오의공학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE