Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Foreign Body Reaction After PLC Reconstruction Caused by a Broken PLLA Screw

Full metadata record
DC Field Value Language
dc.contributor.authorKim, Tae-Kwon-
dc.contributor.authorJeong, Tae-Wan-
dc.contributor.authorLee, Dae-Hee-
dc.date.accessioned2021-09-05T02:32:33Z-
dc.date.available2021-09-05T02:32:33Z-
dc.date.created2021-06-15-
dc.date.issued2014-12-
dc.identifier.issn0147-7447-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/96644-
dc.description.abstractForeign body reactions may occur in patients who receive bioabsorbable implants during orthopedic surgery for fractures and ligament repair. The authors describe a 34-year-old man who presented with a palpable tender mass on the lateral aspect of the left knee of 1 month's duration. He underwent posterior cruciate ligament and posterolateral corner reconstruction 3 years earlier. Physical examination showed a 1x1-cm soft, nontender mass without localized warmth on the lateral epicondyle of the distal femur. Magnetic resonance imaging showed a broken screw fragment surrounded by a cyst-like mass. Under general anesthesia, the surgeon excised the screw fragment and the fibrotic mass, enclosing it in the subcutaneous tissue at the lateral epicondyle, the site at which a poly-L-lactic acid bioabsorbable screw had been inserted to fix the graft for posterolateral corner reconstruction. Histologic evaluation showed a foreign body reaction to the degraded screw particles. To the authors' knowledge, this report is the first description of a patient presenting with a delayed foreign body reaction to a broken poly-L-lactic acid bioabsorbable screw at the lateral femoral epicondyle after posterolateral corner reconstruction. Because delayed foreign body reactions can occur at any site of poly-L-lactic acid bioabsorbable screw insertion, care should be taken to avoid screw protrusion during ligament reconstruction because it can lead to screw breakage and delayed foreign body reaction.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherSLACK INC-
dc.subjectCRUCIATE LIGAMENT RECONSTRUCTION-
dc.subjectINTERFERENCE SCREWS-
dc.subjectCYST FORMATION-
dc.subjectFIXATION-
dc.titleForeign Body Reaction After PLC Reconstruction Caused by a Broken PLLA Screw-
dc.typeArticle-
dc.contributor.affiliatedAuthorLee, Dae-Hee-
dc.identifier.doi10.3928/01477447-20141124-91-
dc.identifier.scopusid2-s2.0-84914821513-
dc.identifier.wosid000347450900013-
dc.identifier.bibliographicCitationORTHOPEDICS, v.37, no.12, pp.E1129 - E1132-
dc.relation.isPartOfORTHOPEDICS-
dc.citation.titleORTHOPEDICS-
dc.citation.volume37-
dc.citation.number12-
dc.citation.startPageE1129-
dc.citation.endPageE1132-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaOrthopedics-
dc.relation.journalWebOfScienceCategoryOrthopedics-
dc.subject.keywordPlusCRUCIATE LIGAMENT RECONSTRUCTION-
dc.subject.keywordPlusINTERFERENCE SCREWS-
dc.subject.keywordPlusCYST FORMATION-
dc.subject.keywordPlusFIXATION-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE