Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Estimate of the shielding effect on secondary cancer risk due to cone-beam CT in image-guided radiotherapy

Full metadata record
DC Field Value Language
dc.contributor.authorSung, Jiwon-
dc.contributor.authorBaek, Tae Seong-
dc.contributor.authorYoon, Myonggeun-
dc.contributor.authorKim, Dong Wook-
dc.contributor.authorKim, Dong Hyun-
dc.date.accessioned2021-09-05T05:51:27Z-
dc.date.available2021-09-05T05:51:27Z-
dc.date.created2021-06-15-
dc.date.issued2014-09-
dc.identifier.issn0374-4884-
dc.identifier.urihttps://scholar.korea.ac.kr/handle/2021.sw.korea/97594-
dc.description.abstractThis study evaluated the effect of a simple shielding method using a thin lead sheet on the imaging dose caused by cone-beam computed tomography (CBCT) in image-guided radiation therapy (IGRT). Reduction of secondary doses from CBCT was measured using a radio-photoluminescence glass dosimeter (RPLGD) placed inside an anthropomorphic phantom. The entire body, except for the region scanned by using CBCT, was shielded by wrapping it with a 2-mm lead sheet. Changes in secondary cancer risk due to shielding were calculated using BEIR VII models. Doses to out-of-field organs for head-and-neck, chest, and pelvis scans were decreased 15 similar to 100%, 23 similar to 90%, and 23 similar to 98%, respectively, and the average reductions in lifetime secondary cancer risk due to the 2-mm lead shielding were 1.6, 11.5, and 12.7 persons per 100,000, respectively. These findings suggest that a simple, thin-lead-sheet-based shielding method can effectively decrease secondary doses to out-of-field regions for CBCT, which reduces the lifetime cancer risk on average by 9 per 100,000 patients.-
dc.languageEnglish-
dc.language.isoen-
dc.publisherKOREAN PHYSICAL SOC-
dc.subjectRADIATION-THERAPY-
dc.subjectGLASS DOSIMETER-
dc.subject2ND CANCERS-
dc.titleEstimate of the shielding effect on secondary cancer risk due to cone-beam CT in image-guided radiotherapy-
dc.typeArticle-
dc.contributor.affiliatedAuthorYoon, Myonggeun-
dc.identifier.doi10.3938/jkps.65.757-
dc.identifier.scopusid2-s2.0-84907660041-
dc.identifier.wosid000342449400028-
dc.identifier.bibliographicCitationJOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.65, no.5, pp.757 - 762-
dc.relation.isPartOfJOURNAL OF THE KOREAN PHYSICAL SOCIETY-
dc.citation.titleJOURNAL OF THE KOREAN PHYSICAL SOCIETY-
dc.citation.volume65-
dc.citation.number5-
dc.citation.startPage757-
dc.citation.endPage762-
dc.type.rimsART-
dc.type.docTypeArticle-
dc.identifier.kciidART001910638-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.description.journalRegisteredClasskci-
dc.relation.journalResearchAreaPhysics-
dc.relation.journalWebOfScienceCategoryPhysics, Multidisciplinary-
dc.subject.keywordPlusRADIATION-THERAPY-
dc.subject.keywordPlusGLASS DOSIMETER-
dc.subject.keywordPlus2ND CANCERS-
dc.subject.keywordAuthorSecondary cancer risk-
dc.subject.keywordAuthorCBCT-
dc.subject.keywordAuthorIGRT-
dc.subject.keywordAuthorBEIR VII-
dc.subject.keywordAuthorShielding-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Myong geun photo

Yoon, Myong geun
보건과학대학 (바이오의공학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE