Monte Carlo simulation-based feasibility study of novel indirect flat panel detector system for removing scatter radiation

  • Yoon, Yongsu
  • Morishita, Junji
  • Park, MinSeok
  • Kim, Hyunji
  • Kim, Kihyun
  • 외 1명
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

3

초록

The purpose of this study is to investigate the feasibility of a novel indirect flat panel detector (FPD) system for removing scatter radiation. The substrate layer of our FPD system has a Pb net-like structure that matches the ineffective area and blocks the scatter radiation such that only primary X-rays reach the effective area on a thin-film transistor. To evaluate the performance of the proposed system, we used Monte Carlo simulations to derive the scatter fraction and contrast. The scatter fraction of the proposed system is lower than that of a parallel grid system, and the contrast is superior to that of a system without a grid. If the structure of the proposed FPD system is optimized with respect to the specifications of a specific detector, the purpose of the examination, and the energy range used, the FPD can be useful in diagnostic radiology. (C) 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

키워드

Indirect flat panel detectorMonte Carlo simulationScatter radiationDIAGNOSTIC-RADIOLOGYDIGITAL RADIOGRAPHYLARGE-AREAPERFORMANCEGRIDS
제목
Monte Carlo simulation-based feasibility study of novel indirect flat panel detector system for removing scatter radiation
저자
Yoon, YongsuMorishita, JunjiPark, MinSeokKim, HyunjiKim, KihyunKim, Jungmin
DOI
10.1016/j.ejmp.2015.11.008
발행일
2016-01
유형
Article
저널명
Physica Medica
32
1
페이지
182 ~ 187