Wide dynamic range X-ray detector utilizing spectroscopic grade thallium bromide single crystal

  • Seo, Jiwon; 
  • Song, Seungho; 
  • Park, Minhwan; 
  • Jo, Ajin; 
  • Lee, Wonho; 
  • 외 5명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

This study aimed to evaluate the radiation response performance of thallium bromide (TlBr) across a wide range of dose rates. The TlBr crystal was purified and grown using a zone-refined purification furnace, and after completing 50 purification cycles, the single crystal was grown with a modified heater translation speed. The properties of the grown crystal were characterized using X-ray diffraction (XRD), current-voltage measurements, photoconductive measurement, and gamma-ray spectroscopy. X-ray performance was assessed under a wide dynamic range of dose rates, X-ray energies, and applied bias voltages. The results demonstrated a minimum detectable limit (MDL) of 130.19 nGy/s, confirming the detector's capability to respond to extremely low-dose Xray exposure. Owing to its excellent material properties, X-ray measurements exhibited a high signal-to-noise ratio, enabling high-contrast X-ray imaging.

키워드

Compound semiconductor; Thallium bromide; X-ray imaging; Minimum detection limit; Radiation detection; CDTE
제목
Wide dynamic range X-ray detector utilizing spectroscopic grade thallium bromide single crystal
저자
Seo, Jiwon; Song, Seungho; Park, Minhwan; Jo, Ajin; Lee, Wonho; Kim, Eunhye; Kim, Younghak; Jang, Mee; Park, Beomjun; Yeom, Jung-Yeol
DOI
10.1016/j.net.2025.103817
발행일
2025-12
유형
Article
저널명
Nuclear Engineering and Technology
권
57
호
12