Establishing Cutoff Values for a Quality Assurance Test Using an Ultrasound Phantom in Screening Ultrasound Examinations for Hepatocellular Carcinoma An Initial Report of a Nationwide Survey in Korea

  • Choi, Joon-Il
  • Kim, Pyo Nyun
  • Jeong, Woo Kyoung
  • Kim, Hyun Cheol
  • Yang, Dal Mo
  • 외 2명
Citations

WEB OF SCIENCE

9
Citations

SCOPUS

12

초록

Objectives-The purpose of this study was to evaluate the results of ultrasound (US) device testing using a US phantom and to determine cutoff values for phantom quality assurance tests in US examinations for the screening of hepatocellular carcinoma in Korea. Methods-Ultrasound phantom images were acquired from the general hospitals in Korea that participated in the National Cancer Screening Program for hepatocellular carcinoma. Ultrasound images of the phantom were acquired with a 3.0- to 5.0-MHz convex transducer and evaluated in terms of the dead zone, vertical and horizontal measurement, axial and lateral resolution, sensitivity, and gray scale/dynamic range. Appropriate cutoff values were determined to guarantee minimal qualifications for the performance of the US scanners. Results-Three hundred fifty-seven US scanners were tested using the following cutoff values: less than 2 mm for the dead zone, 5% discrepancy in the vertical measurement, 7.5% discrepancy in the horizontal measurement, all I 1 identifiable line targets for axial and lateral resolution, more than 14 cm for sensitivity, and more than 4 cylindrical structures for gray scale/dynamic range. With these criteria, 283 US scanners (79.3%) passed the tests. The most common cause of disqualification was the dynamic range/gray scale. No statistical difference was observed in the disqualification rate between 3 groups based on different years of manufacture. Conclusions-Through this study, we have defined cutoff values for phantom images acquired with US scanners. These will be used in performing screening US examinations for hepatocellular carcinoma in Korea.

키워드

hepatocellular carcinomaphantomquality assuranceultrasoundHIGH AGREEMENTLOW KAPPAMANAGEMENT
제목
Establishing Cutoff Values for a Quality Assurance Test Using an Ultrasound Phantom in Screening Ultrasound Examinations for Hepatocellular Carcinoma An Initial Report of a Nationwide Survey in Korea
저자
Choi, Joon-IlKim, Pyo NyunJeong, Woo KyoungKim, Hyun CheolYang, Dal MoCha, Sang HoonChung, Jae-Joon
발행일
2011-09
유형
Article
저널명
Journal of Ultrasound in Medicine
30
9
페이지
1221 ~ 1229