Applicability of Glass Dosimeters for In-vivo Dosimetry in Brachytherapy

  • Moon, Sun Young; 
  • Son, Jaeman; 
  • Yoon, Myonggeun; 
  • Jeang, EunHee; 
  • Lim, Young Kyung; 
  • 외 2명
Citations

WEB OF SCIENCE

3
Citations

SCOPUS

2

초록

During brachytherapy, confirming the dose delivered is very important in order to prevent radiation-associated side effects. Therefore, we aimed to confirm the accuracy of dose delivery near the source by inserting glass dosimeters within the applicator. We created an alternative pelvic phantom with the same shape and internal structures as the usual patient. In addition, we created a tandem for insertion of the glass dosimeters and measured the dose near the source by inserting the glass dosimeters into the tandem and evaluating the accuracy of the dwell position and time through the dose near the source. Errors between the values obtained from the five glass dosimeters and the values from the treatment planning system were -6.27, -2.1, -4.18, 6.31, and -0.39%, respectively. The mean error was 3.85%. This value was acceptable considering that the error of the glass dosimeter itself is approximately 3%. Even though a complement of the applicator and the error calibration is required in order to apply this technique clinically, we believe that radiation accidents and overdoses can be prevented through in-vivo dosimetry using a glass dosimeter for brachytherapy.

키워드

Brachytherapy; Glass dosimeter; In-vivo dosimetry; Patient dose; DOSE-RATE BRACHYTHERAPY; IR-192 HDR BRACHYTHERAPY; THERMOLUMINESCENCE DOSIMETRY; SCINTILLATION DETECTORS; PROSTATE-CANCER; ROD DETECTOR; VERIFICATION; PHANTOM; ARRAY
제목
Applicability of Glass Dosimeters for In-vivo Dosimetry in Brachytherapy
저자
Moon, Sun Young; Son, Jaeman; Yoon, Myonggeun; Jeang, EunHee; Lim, Young Kyung; Chung, Weon Kyu; Kim, Dong Wook
DOI
10.3938/jkps.72.1320
발행일
2018-06
유형
Article
저널명
Journal of the Korean Physical Society
권
72
호
11
페이지
1320 ~ 1325