Development of beam monitoring system for proton pencil beam scanning using fiber-optic radiation sensor

Citations

WEB OF SCIENCE

5
Citations

SCOPUS

6

초록

We aimed to develop a beam monitoring system based on a fiber-optic radiation sensor (FORS), which can be used in real time in a beam control room, to monitor a beam in proton therapy, where patients are treated using a pencil beam scanning (PBS) mode, by measuring the beam spot width (BSW) and beam spot position (BSP) of the PBS. We developed two-dimensional detector arrays to monitor the PBS beam in the beam control room. We measured the BSW for five energies of the PBS beam and compared the measurements with those of Lynx and EBT3 film. In order to confirm the BSP, we compared the BSP values of the PBS calculated from radiation treatment planning (RTP), to five BSP values measured using FORS at 224.2 MeV. When comparing BSW values obtained using developed monitoring system to the measurements obtained using commercial EBT3 film, the average difference in BSW value of the PBS beam was 0.1 +/- 0.1 mm. In the comparison of BSW values with the measurements obtained using Lynx, the average difference was 0.2 +/- 0.1 mm. When comparing BSP measurements to the values calculated from RTP, the average difference was 0.4 +/- 0.2 mm. The study results confirmed that the developed FORS-based beam monitoring system can monitor a PBS beam in real time in a beam control room, where proton beam is controlled for the patient.

키워드

Proton therapy; Beam monitoring system; Fiber-optic radiation sensor; THERAPY; VERIFICATION; IMRT
제목
Development of beam monitoring system for proton pencil beam scanning using fiber-optic radiation sensor
저자
Son, Jaeman; Koo, Jihye; Moon, Sunyoung; Yoon, Myonggeun; Jeong, Jonghwi; Kim, Sun-Young; Lim, Youngkyung; Lee, Se Byeong; Shin, Dongho; Kim, Meyoung; Kim, Dongwook
DOI
10.3938/jkps.71.438
발행일
2017-10
유형
Article
저널명
Journal of the Korean Physical Society
권
71
호
7
페이지
438 ~ 443