Evaluation of mechanical accuracy for couch-based tracking system (CBTS)

Citations

WEB OF SCIENCE

11
Citations

SCOPUS

10

초록

This study evaluated the mechanical accuracy of an in-house-developed couch-based tracking system (CBTS) according to respiration data. The overall delay time of the CBTS was calculated, and the accuracy, reproducibility, and loading effect of the CBTS were evaluated according to the sinusoidal waveform and various respiratory motion data of real patients with and without a volunteer weighing 75kg. The root mean square (rms) error of the accuracy, the reproducibility, and the sagging measurements were calculated for the three axes (X, Y, and Z directions) of the CBTS. The overall delay time of the CBTS was 0.251 sec. The accuracy and reproducibility in the Z direction in real patient data were poor, as indicated by high rms errors. The results of the loading effect were within 1.0 mm in all directions. This novel CBTS has the potential for clinical application for tumor tracking in radiation therapy.

키워드

couchtrackingimage-guided radiation therapy (IGRT)delay timeconsistencyBREATH-HOLD TECHNIQUEIMAGE-GUIDED RADIOTHERAPYRESPIRATORY TUMOR MOTIONREAL-TIME TRACKINGDOSE-ESCALATIONPROSTATE-CANCERTARGET TRACKINGLUNGPREDICTIONMANAGEMENT
제목
Evaluation of mechanical accuracy for couch-based tracking system (CBTS)
저자
Lee, SukChang, Kyung-HwanShim, Jand BoCao, YuanjieLee, Chang KiCho, Sam JuYang, Dae SikPark, Young JeYoon, Won SeobKim, Chul Yong
DOI
10.1120/jacmp.v13i6.3818
발행일
2012
유형
Article
저널명
Journal of Applied Clinical Medical Physics
13
6
페이지
157 ~ 169