Detection of IMRT delivery errors based on a simple constancy check of transit dose by using an EPID

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

Beam delivery errors during intensity modulated radiotherapy (IMRT) were detected based on a simple constancy check of the transit dose by using an electronic portal imaging device (EPID). Twenty-one IMRT plans were selected from various treatment sites, and the transit doses during treatment were measured by using an EPID. Transit doses were measured 11 times for each course of treatment, and the constancy check was based on gamma index (3%/3 mm) comparisons between a reference dose map (the first measured transit dose) and test dose maps (the following ten measured dose maps). In a simulation using an anthropomorphic phantom, the average passing rate of the tested transit dose was 100% for three representative treatment sites (head & neck, chest, and pelvis), indicating that IMRT was highly constant for normal beam delivery. The average passing rate of the transit dose for 1224 IMRT fields from 21 actual patients was 97.6% +/- 2.5%, with the lower rate possibly being due to inaccuracies of patient positioning or anatomic changes. An EPIDbased simple constancy check may provide information about IMRT beam delivery errors during treatment.

키워드

Transit dose; In vivo dosimetry; EPID; Intensity modulated radiation therapy; PRETREATMENT VERIFICATION; CLINICAL-EXPERIENCE; PORTAL DOSIMETRY; DYNAMIC IMRT; DIODE-ARRAY
제목
Detection of IMRT delivery errors based on a simple constancy check of transit dose by using an EPID
저자
Baek, Tae Seong; Chung, Eun Ji; Son, Jaeman; Yoon, Myonggeun
DOI
10.3938/jkps.67.1876
발행일
2015-11
유형
Article
저널명
Journal of the Korean Physical Society
권
67
호
10
페이지
1876 ~ 1881