LED and CMOS image sensor based hemoglobin concentration measurement technique

Citations

WEB OF SCIENCE

25
Citations

SCOPUS

28

초록

This paper presents a technique that can provide accurate hemoglobin concentration measurement within a simple, cost-effective and compact platform. It is based on a LED and CMOS image sensor without using an environmentally harmful reagent and centrifugal process. The acquired standard curve and adequate image processing steps such as image calibration and noise suppression allow us to accurately estimate the hemoglobin concentration at an error rate of less than 2% for blind tests conducted with the 33 blood specimens of outpatients. The hemoglobin concentration measurement, one of the crucial part of the complete blood count (CBC) as well as the most commonly ordered test in clinical laboratory, within such a small, cheap and fast platform would be useful for wide applications in doctor's office and for applications in resource limited settings as a point-of-care testing tool. (C) 2011 Elsevier B.V. All rights reserved.

키워드

Hemoglobin concentration measurementLEDCMOS image sensorSpectrophotometryBlood analysisCHIPSTANDARDIZATION
제목
LED and CMOS image sensor based hemoglobin concentration measurement technique
저자
Kim, Dong-SikChoi, Jae-HoonNam, Myung-HyunYang, Ji-WoonPak, James JunghoSeo, Sungkyu
DOI
10.1016/j.snb.2011.03.032
발행일
2011-09-20
유형
Article
저널명
Sensors and Actuators, B: Chemical
157
1
페이지
103 ~ 109