Effect of topogram-tube angle combination on CT radiation dose reduction

Citations

WEB OF SCIENCE

3
Citations

SCOPUS

3

초록

This study assessed the ability of various types of topograms, when used with an automatic tube current modulation (ATCM) technique, to reduce radiation dose from computed tomography (CT) scans. Three types of topograms were used with the ATCM technique: (i) anteroposterior (AP) topograms alone, (ii) AP topograms followed by lateral topograms, and (iii) lateral topograms followed by AP topograms. Various regions (chest, abdomen and whole-body) of a humanoid phantom were scanned at several tube voltages (80, 100 and 120 kVp) with the selected topograms. Although the CT dose depended on the order of topograms, the CT dose with respect to patient positioning depended on the number of topograms performed. The magnitude of the difference in CT dose between number and order of topograms was greater for the scans of the abdomen than the chest. These results suggest that, for the Siemens SOMATOM Definition AS CT scanner, choosing the right combination of CT scan conditions with the ATCM technique can minimize radiation dose to a patient.

키워드

Computerized Tomography (CT) and Computed Radiography (CR); Radiation calculations; AUTOMATIC EXPOSURE CONTROL; CURRENT MODULATION; COMPUTED-TOMOGRAPHY; IMAGE QUALITY; PATIENT; PHANTOM; NOISE
제목
Effect of topogram-tube angle combination on CT radiation dose reduction
저자
Shim, J.; Yoon, M.
DOI
10.1088/1748-0221/12/09/T09003
발행일
2017-09
유형
Article
저널명
Journal of Instrumentation
권
12