Nano-Interstice Driven Powerless Blood Plasma Extraction in a Membrane Filter Integrated Microfluidic Device

  • Kim, Jaehoon
  • Yoon, Junghyo
  • Byun, Jae-Yeong
  • Kim, Hyunho
  • Han, Sewoon
  • ... Chung, Seok
  • 외 3명
Citations

WEB OF SCIENCE

10
Citations

SCOPUS

11

초록

Blood plasma is a source of biomarkers in blood and a simple, fast, and easy extraction method is highly required for point-of-care testing (POCT) applications. This paper proposes a membrane filter integrated microfluidic device to extract blood plasma from whole blood, without any external instrumentation. A commercially available membrane filter was integrated with a newly designed dual-cover microfluidic device to avoid leakage of the extracted plasma and remaining blood cells. Nano-interstices installed on both sides of the microfluidic channels actively draw the extracted plasma from the membrane. The developed device successfully supplied 20 mu L of extracted plasma with a high extraction yield (similar to 45%) in 16 min.

키워드

microfluidicspoint-of-care testingblood plasma extraction
제목
Nano-Interstice Driven Powerless Blood Plasma Extraction in a Membrane Filter Integrated Microfluidic Device
저자
Kim, JaehoonYoon, JunghyoByun, Jae-YeongKim, HyunhoHan, SewoonKim, JunghyunLee, Jeong HoonJo, Han-SangChung, Seok
DOI
10.3390/s21041366
발행일
2021-02
유형
Article
저널명
Sensors
21
4
페이지
1 ~ 19