Feasibility Study of a No-Insulation 1.5-T/600-mm All-REBCO Magnet for MRI Systems
- Kim, Kwang Lok;
- Song, Jung-Bin;
- Choi, Yoon Hyuck;
- Yang, Dong Gyu;
- Lee, Haigun
WEB OF SCIENCE
22SCOPUS
22초록
This paper presents the design of a 1.5-T/600-mm all-REBCO magnetic resonance imaging (MRI) magnet comprising a stack of 118 no-insulation (NI) double-pancake coils. A 4.1-mm-wide and 0.1-mm-thick REBCO tape was used in this study. The preshim overall field homogeneity within a 50-cm diameter of spherical volume was estimated to be <25 ppm. According to the equivalent-circuit model of the NI magnet, the basic magnet performance was demonstrated analytically. The design results of the 1.5-T/600-mm all-REBCO MRI magnet with the NI technology indicate the possibility that the NI winding technology can be used in all-REBCO NI magnets for MRI systems. Consequently, the combination of the REBCO tape and the NI winding technique is expected to enable the development of 1.5-T MRI magnets that offer LHe-free operation and are self-protecting, highly compact, and mechanically robust.
키워드
- 제목
- Feasibility Study of a No-Insulation 1.5-T/600-mm All-REBCO Magnet for MRI Systems
- 저자
- Kim, Kwang Lok; Song, Jung-Bin; Choi, Yoon Hyuck; Yang, Dong Gyu; Lee, Haigun
- 발행일
- 2017-06
- 유형
- Article
- 권
- 27
- 호
- 4
- 언어
- ENG
- 출판사
- IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
- 발행국가
- 미국
- ISSN
- E 1558-2515