Feasibility Study of a No-Insulation 1.5-T/600-mm All-REBCO Magnet for MRI Systems

  • Kim, Kwang Lok; 
  • Song, Jung-Bin; 
  • Choi, Yoon Hyuck; 
  • Yang, Dong Gyu; 
  • Lee, Haigun
Citations

WEB OF SCIENCE

22
Citations

SCOPUS

22

초록

This paper presents the design of a 1.5-T/600-mm all-REBCO magnetic resonance imaging (MRI) magnet comprising a stack of 118 no-insulation (NI) double-pancake coils. A 4.1-mm-wide and 0.1-mm-thick REBCO tape was used in this study. The preshim overall field homogeneity within a 50-cm diameter of spherical volume was estimated to be <25 ppm. According to the equivalent-circuit model of the NI magnet, the basic magnet performance was demonstrated analytically. The design results of the 1.5-T/600-mm all-REBCO MRI magnet with the NI technology indicate the possibility that the NI winding technology can be used in all-REBCO NI magnets for MRI systems. Consequently, the combination of the REBCO tape and the NI winding technique is expected to enable the development of 1.5-T MRI magnets that offer LHe-free operation and are self-protecting, highly compact, and mechanically robust.

키워드

Charging delay; MRI magnets; no-insulation; REBCO tape
제목
Feasibility Study of a No-Insulation 1.5-T/600-mm All-REBCO Magnet for MRI Systems
저자
Kim, Kwang Lok; Song, Jung-Bin; Choi, Yoon Hyuck; Yang, Dong Gyu; Lee, Haigun
DOI
10.1109/TASC.2016.2628880
발행일
2017-06
유형
Article
저널명
IEEE Transactions on Applied Superconductivity
권
27
호
4