Separation of platelets from whole blood using standing surface acoustic waves in a microchannel
- Nam, Jeonghun;
- Lim, Hyunjung;
- Kim, Dookon;
- Shin, Sehyun
WEB OF SCIENCE
162SCOPUS
174초록
Platelet separation from blood is essential for biochemical analyses and clinical diagnosis. In this article, we propose a method to separate platelets from undiluted whole blood using standing surface acoustic waves (SSAWs) in a microfluidic device. A polydimethylsiloxane (PDMS) microfluidic channel was fabricated and integrated with interdigitated transducer (IDT) electrodes patterned on a piezoelectric substrate. To avoid shear-induced activation of platelets, the blood sample flow was hydrodynamically focused by introducing sheath flow from two side-inlets and pressure nodes were designed to locate at side walls. By means of flow cytometric analysis, the RBC clearance ratio from whole blood was found to be over 99% and the purity of platelets was close to 98%. Conclusively, the present technique using SSAWs can directly separate platelets from undiluted whole blood with higher purity than other methods.
키워드
- 제목
- Separation of platelets from whole blood using standing surface acoustic waves in a microchannel
- 저자
- Nam, Jeonghun; Lim, Hyunjung; Kim, Dookon; Shin, Sehyun
- 발행일
- 2011
- 유형
- Article
- 저널명
- Lab on a Chip
- 권
- 11
- 호
- 19
- 페이지
- 3361 ~ 3364