A cubic gamma camera with an active collimator

Citations

WEB OF SCIENCE

9
Citations

SCOPUS

9

초록

Mechanical collimation with photon absorption and electronic collimation using Compton scattering are combined to form a cubic gamma camera with an active collimator. The collimator is made active by constructing it with a uniformly redundant array of patterned Bi4Ge3O12 (BGO) scintillators, which not only attenuates incident radiations but also detects scattered radiation, in a gamma-camera consisting of and five planar CsI(Na) scintillators. The entire module forms a cubic structure that generates images on the basis of radiation interactions from every direction. The coverage angle for detecting scattered radiation is 2 pi with a detection efficiency approximately 17 times higher than previous systems that comprised only one pair of detectors. (C) 2014 Elsevier Ltd. All rights reserved.

키워드

Cubic gamma cameraMechanical collimationElectronic collimationActive collimatorEMISSION COMPUTED-TOMOGRAPHYPHOTOMULTIPLIER TUBESMAXIMUM-LIKELIHOODRECONSTRUCTIONSCINTILLATORSARRAYSIMAGER
제목
A cubic gamma camera with an active collimator
저자
Lee, TaewoongLee, Wonho
DOI
10.1016/j.apradiso.2014.03.020
발행일
2014-08
유형
Article
저널명
Applied Radiation and Isotopes
90
페이지
102 ~ 108