A cubic gamma camera with an active collimator

Citations

WEB OF SCIENCE

9
Citations

SCOPUS

9

초록

Mechanical collimation with photon absorption and electronic collimation using Compton scattering are combined to form a cubic gamma camera with an active collimator. The collimator is made active by constructing it with a uniformly redundant array of patterned Bi4Ge3O12 (BGO) scintillators, which not only attenuates incident radiations but also detects scattered radiation, in a gamma-camera consisting of and five planar CsI(Na) scintillators. The entire module forms a cubic structure that generates images on the basis of radiation interactions from every direction. The coverage angle for detecting scattered radiation is 2 pi with a detection efficiency approximately 17 times higher than previous systems that comprised only one pair of detectors. (C) 2014 Elsevier Ltd. All rights reserved.

키워드

Cubic gamma camera; Mechanical collimation; Electronic collimation; Active collimator; EMISSION COMPUTED-TOMOGRAPHY; PHOTOMULTIPLIER TUBES; MAXIMUM-LIKELIHOOD; RECONSTRUCTION; SCINTILLATORS; ARRAYS; IMAGER
제목
A cubic gamma camera with an active collimator
저자
Lee, Taewoong; Lee, Wonho
DOI
10.1016/j.apradiso.2014.03.020
발행일
2014-08
유형
Article
저널명
Applied Radiation and Isotopes
권
90
페이지
102 ~ 108