We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired t-test, p < 0.05). At the last follow-up, the mean maximum flexion was 135° (110° to 150°). Two knees showed radiolucent lines, but revision was not required because the patients were asymptomatic. Revision was required in one case because of infection, but there were no prosthesis-related revisions. There were no other complications. The estimated survival rate at four years with revision for any reason and prosthesis-related problems was 99.6% and 100%, respectively. This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant. ©2008 British Editorial Society of Bone and Joint Surgery.

키워드

adultagedarticlecontrolled studyfemalefollow uphumaninfectionknee prosthesismajor clinical studymaleoutcome assessmentpostoperative complicationpriority journalprospective studyprosthesis failuresurvivalsurvival ratetotal knee replacementAgedAged, 80 and overArthroplasty, Replacement, KneeFemaleFollow-Up StudiesHumansKnee JointKnee ProsthesisMaleMiddle AgedOsteoarthritis, KneePostoperative ComplicationsProspective StudiesProsthesis DesignRange of Motion, ArticularTreatment Outcome
제목
The NexGen LPS-flex to the knee prosthesis at a minimum of three years
저자
Kim, T.-H.Lee, D.-H.Bin, S.-I.
DOI
10.1302/0301-620X.90B10.21050
발행일
2008
유형
Article
저널명
Journal of Bone and Joint Surgery - Series B
90
10
페이지
1304 ~ 1310