Oxford phase III meniscal bearing fracture: Case report

  • Lim, Hong-Chul
  • Shon, Won-Yong
  • Kim, Seung-Ju
  • Bae, Ji-Hoon
Citations

WEB OF SCIENCE

9
Citations

SCOPUS

9

초록

Meniscal bearing fracture is a rare complication of phase III Oxford unicompartmental knee replacement (UKR). We report a case of a meniscal bearing fracture that occurred 7 years after phase III Oxford medial UKR The meniscal bearing showed uneven delamination of the polyethylene in the thinnest articular surface and an impingement lesion. This lesion initiated a fatigue crack that propagated to cause failure of the meniscal bearing. This is the first report of a meniscal bearing fracture without a posterior marker wire. (C) 2013 Elsevier B.V. All rights reserved.

키워드

Oxford unicompartmentalKneeMeniscal bearingFractureUNICOMPARTMENTAL KNEE REPLACEMENTPOLYETHYLENE WEARARTHROPLASTYRETRIEVAL
제목
Oxford phase III meniscal bearing fracture: Case report
저자
Lim, Hong-ChulShon, Won-YongKim, Seung-JuBae, Ji-Hoon
DOI
10.1016/j.knee.2013.07.012
발행일
2014-01
유형
Article
저널명
The Knee
21
1
페이지
340 ~ 342