상세 보기
초록
High stiffness frame design for a spine manipulation device was developed in this research. For the safety of a spinal manipulation, high stiffness of the device is required. A finite element (FE)model of the device frame is created and validated by measured vibration data. Parameters are suggested for high stiffness design of the frame. Based on the Taguchi design of experiment (DOE), a practical set of design parameter values is suggested.
키워드
Stiffness (강성); Finite Element (유한요소); Design of Experiment (실험 계획법); Spine Manipulation Device (척추교정 장치)
- 제목
- 척추교정 장치의 고강성 프레임 설계
- 제목 (타언어)
- High Stiffness Frame Design for a Spine Manipulation Device
- 저자
- 문영환; 김정훈; 김권희
- 발행일
- 2011
- 저널명
- 한국정밀공학회지
- 권
- 28
- 호
- 1
- 페이지
- 89 ~ 95