치과용 백색광 스캐너를 이용한 impression scanning의 반복 측정에 대한 안정성 평가

Evaluation on the repeatability of dental white light scanner-based digital impression

초록

Purpose: The purpose of this study was to evaluate the repeatability of dental white light scanner. Methods: The impression(Zerosil, Dreve, Germany) were digitized in white light scanner(Identica, Medit, Korea)to create 3-dimensional surface-models. The distribution of the discrepancies between the number of points in the corresponding CRM models and the point clouds in the others were measured by a matching-software(Power-Inspect 2012, Delcam Plc, UK). The discriptive statistics were used for statistical analysis(SPSS 20.0). Results: The measurement of repeatablity showed very good reliability. The mean(SD) discrepancy value on the white light scanner digital models was 8.7(0.67) , based on SD and absolute mean values. Conclusion: These in vitro studies showed that repeatability of dental white light scanner is high reliability. These results can be confirmed in further clinical studies.

키워드

dental white light scannerrepeatability of dental scannerdigital impressionCRM modelspoint clouds