치과용 백색광 스캐너를 이용한 impression scanning의 반복 측정에 대한 안정성 평가
- 전진훈;
- 김웅철;
- 이경탁;
- 김혜영;
- 김지환
초록
Purpose: The purpose of this study was to evaluate the repeatability of dental white light scanner. Methods: The impression(Zerosil, Dreve, Germany) were digitized in white light scanner(Identica, Medit, Korea)to create 3-dimensional surface-models. The distribution of the discrepancies between the number of points in the corresponding CRM models and the point clouds in the others were measured by a matching-software(Power-Inspect 2012, Delcam Plc, UK). The discriptive statistics were used for statistical analysis(SPSS 20.0). Results: The measurement of repeatablity showed very good reliability. The mean(SD) discrepancy value on the white light scanner digital models was 8.7(0.67) , based on SD and absolute mean values. Conclusion: These in vitro studies showed that repeatability of dental white light scanner is high reliability. These results can be confirmed in further clinical studies.
키워드
- 제목
- 치과용 백색광 스캐너를 이용한 impression scanning의 반복 측정에 대한 안정성 평가
- 제목 (타언어)
- Evaluation on the repeatability of dental white light scanner-based digital impression
- 저자
- 전진훈; 김웅철; 이경탁; 김혜영; 김지환
- 발행일
- 2013
- 저널명
- 대한치과기공학회지
- 권
- 35
- 호
- 1
- 페이지
- 37 ~ 42