DOSE EVALUATION OF SELECTIVE COLLIMATION EFFECT IN CEPHALOGRAPHY BY MEASUREMENT AND MONTE CARLO SIMULATION

  • Lee, Boram
  • Shin, Gwisoon
  • Kang, Sunjung
  • Shin, Boram
  • Back, Ilhong
  • ... Lee, Wonho
  • 외 6명
Citations

WEB OF SCIENCE

13
Citations

SCOPUS

12

초록

Recently, simulations based on the Monte Carlo code have been increasingly applied for physics phenomena, patient dose and quality assurance of radiation systems. The objective of this study was to use Monte Carlo simulation and measurement to verify dose and dose reduction in cephalography. The collimator was constructed with 3-mm thick lead plate, and attached to the tube head to remove regions of disinterest in the radiation field. A digital phantom patient was constructed to evaluate patient dose. In addition, detectors of pixel size 1 x 1 cm(2) and 0.1 x 0.1 cm(2) were constructed to check collimator location. The effective dose according to International Commission on Radiological Protection 103 was calculated with and without collimation. The effective doses for simulation with and without collimation were 5.09 and 11.32 mu Sv, respectively. The results of the calculated effective dose show 61.7% reduction of field area and 55%, of effective dose. The Monte Carlo simulation is a good evaluation tool for patient dose.

제목
DOSE EVALUATION OF SELECTIVE COLLIMATION EFFECT IN CEPHALOGRAPHY BY MEASUREMENT AND MONTE CARLO SIMULATION
저자
Lee, BoramShin, GwisoonKang, SunjungShin, BoramBack, IlhongPark, HyokPark, ChangseoLee, JeongwooLee, WonhoChoi, JonghakPark, RyeonghwangKim, Youhyun
DOI
10.1093/rpd/ncq596
발행일
2012
유형
Article
저널명
Radiation Protection Dosimetry
148
1
페이지
58 ~ 64