심부체온의 감시 오류로 발생한 저체온증 -증례보고-

Hypothermia induced by Incorrect Monitoring of Core Temperature -A case report-
  • 반영준
  • 신혜원
  • 이혜원
  • 임혜자
  • 윤석민
  • 외 1명

초록

A 55-year-old man was scheduled to undergo revisional total hip arthroplasty under general anesthesia. During the operation, a warm blanket and fluid warmer were applied, with his body temperature maintained at 35oC, as assessed with the use of a temporal-artery thermometer. In the course of the operation, severe bleeding developed, which required a massive transfusion of packed red blood cells and other blood components to correct the electrolyte and pH balances; the infusion of cardiovascular drugs was also preformed. However, the hemodynamic status of the patient deteriorated to severe hypotension, with atrial fibrillation, non-sustaining paroxysmal ventricular tachycardia and ventricular fibrillation. The patient was re-evaluated, and it was realized the temporal-artery thermometer had been incorrectly monitored, causing hemodynamic deterioration, which was assessed as hypothermia from a nasopharyngeal temperature of 29oC. Active warming methods, including a condensed humidifier, and warming of the fluids and blood compounds with a rapid infusion system, were instigated, resulting in stabilization of the patient's hemodynamic status and the disappearance of his dysrhythmia. (Korean J Anesthesiol 2006; 50: 217~20)

키워드

hypothermiatemperature monitoringtemporal-artery thermometer.hypothermiatemperature monitoringtemporal-artery thermometer.
제목
심부체온의 감시 오류로 발생한 저체온증 -증례보고-
제목 (타언어)
Hypothermia induced by Incorrect Monitoring of Core Temperature -A case report-
저자
반영준신혜원이혜원임혜자윤석민장성호
발행일
2006
저널명
Korean Journal of Anesthesiology
50
2
페이지
217 ~ 220