상세 보기
초록
A 55-year-old man was scheduled to undergo revisional total hip arthroplasty under general anesthesia. During the operation, a warm blanket and fluid warmer were applied, with his body temperature maintained at 35oC, as assessed with the use of a temporal-artery thermometer. In the course of the operation, severe bleeding developed, which required a massive transfusion of packed red blood cells and other blood components to correct the electrolyte and pH balances; the infusion of cardiovascular drugs was also preformed. However, the hemodynamic status of the patient deteriorated to severe hypotension, with atrial fibrillation, non-sustaining paroxysmal ventricular tachycardia and ventricular fibrillation. The patient was re-evaluated, and it was realized the temporal-artery thermometer had been incorrectly monitored, causing hemodynamic deterioration, which was assessed as hypothermia from a nasopharyngeal temperature of 29oC. Active warming methods, including a condensed humidifier, and warming of the fluids and blood compounds with a rapid infusion system, were instigated, resulting in stabilization of the patient's hemodynamic status and the disappearance of his dysrhythmia. (Korean J Anesthesiol 2006; 50: 217~20)
키워드
- 제목
- 심부체온의 감시 오류로 발생한 저체온증 -증례보고-
- 제목 (타언어)
- Hypothermia induced by Incorrect Monitoring of Core Temperature -A case report-
- 저자
- 반영준; 신혜원; 이혜원; 임혜자; 윤석민; 장성호
- 발행일
- 2006
- 권
- 50
- 호
- 2
- 페이지
- 217 ~ 220