Control of multi-dof ultrasonic actuator for dexterous surgical instrument

Citations

WEB OF SCIENCE

39
Citations

SCOPUS

49

초록

For surgical devices used in minimally invasive surgery, compact multi-degree-of-freedom (multi-dof) actuators are required due to a small work space in the patient body. With conventional single-dof actuators such as electromagnetic motors, a multiple number of actuators are required to generate multi-dof motion. Moreover, additional transmission mechanisms make the total system even larger and heavier. In contrast, a single unit of multi-dof ultrasonic actuator can generate multi-dof rotation of a spherical rotor by utilizing three natural vibration modes of a bar-shaped stator. In our previous work we designed and built a surgical forceps with a wrist capable of two-dof motion, using a single multi-dof ultrasonic motor. The present study proposes a novel control scheme to generate three-dof motion of the surgical instrument for minimally invasive surgical procedures. Through experiments we have confirmed the system performance of the developed device. (c) 2007 Elsevier Ltd. All rights reserved.

제목
Control of multi-dof ultrasonic actuator for dexterous surgical instrument
저자
Takemura, KenjiroPark, ShinsukMaeno, Takashi
DOI
10.1016/j.jsv.2007.09.040
발행일
2008-04-08
유형
Article
저널명
Journal of Sound and Vibration
311
3-5
페이지
652 ~ 666