Influence of an active vibration isolator and electrochemical cell design on electrochemical measurements to minimize natural convection

Citations

WEB OF SCIENCE

4
Citations

SCOPUS

4

초록

The influence of an active vibration isolator (AVI) and electrochemical cell design on natural convection was investigated. The natural convection of an electrolyte caused by external vibrations and acoustic sounds is reduced by both an AVI and an acoustic enclosure. Slow scan voltammetry with an AVI shows a reproducible and predictable diffusion-controlled current down to 0.4 mV/s. Reducing the distance between the working electrode and the glass wall also has a considerable effect on the voltammetric signal.

키워드

Natural convectionSlow scan voltammetryActive vibration isolationAcoustic enclosureElectrochemical cell designMASS-TRANSPORTVOLTAMMETRYELECTRODETEMPERATURECURRENTS
제목
Influence of an active vibration isolator and electrochemical cell design on electrochemical measurements to minimize natural convection
저자
Park, KyungsoonKim, EunsungPark, Jun HuiHwang, Seongpil
DOI
10.1016/j.elecom.2017.07.017
발행일
2017-09
유형
Article
저널명
Electrochemistry Communications
82
페이지
93 ~ 97