상세 보기
Influence of an active vibration isolator and electrochemical cell design on electrochemical measurements to minimize natural convection
- Park, Kyungsoon;
- Kim, Eunsung;
- Park, Jun Hui;
- Hwang, Seongpil
Citations
WEB OF SCIENCE
4Citations
SCOPUS
4초록
The influence of an active vibration isolator (AVI) and electrochemical cell design on natural convection was investigated. The natural convection of an electrolyte caused by external vibrations and acoustic sounds is reduced by both an AVI and an acoustic enclosure. Slow scan voltammetry with an AVI shows a reproducible and predictable diffusion-controlled current down to 0.4 mV/s. Reducing the distance between the working electrode and the glass wall also has a considerable effect on the voltammetric signal.
키워드
Natural convection; Slow scan voltammetry; Active vibration isolation; Acoustic enclosure; Electrochemical cell design; MASS-TRANSPORT; VOLTAMMETRY; ELECTRODE; TEMPERATURE; CURRENTS
- 제목
- Influence of an active vibration isolator and electrochemical cell design on electrochemical measurements to minimize natural convection
- 저자
- Park, Kyungsoon; Kim, Eunsung; Park, Jun Hui; Hwang, Seongpil
- 발행일
- 2017-09
- 유형
- Article
- 권
- 82
- 페이지
- 93 ~ 97