초록
A hydrogen purifier equipped with Pd-Cu-Ni/PNS membranes has been developed to purify low-grade hydrogen and supply it to processes requiring high-purity hydrogen. The purifier does not include any purge system to flush out hydrogen from the membrane module to prevent membrane embrittlement because there is no alpha-beta phase transition below the critical point of the Pd-H system, making the purifier simple. The hydrogen purifier was tested with three different grades of hydrogen, 90, 99 and 99.9%, to determine the effects of the grade of feed hydrogen on the hydrogen permeation behavior. A lower grade required a lower recovery ratio of the purifier to obtain a given relative hydrogen permeation flux. It was confirmed that the purifier can provide high-purity hydrogen to a gas chromatograph (GC) for carrier and make-up gases. A 75-day durability test provided evidence that the hydrogen purifier could be useful for extended periods as needed for commercial processes.
키워드
- 제목
- Development of a hydrogen purifier with Pd-based composite membrane
- 저자
- Ryi, Shin-Kun; Park, Jong-Soo; Kim, Sung-Hyun; Kim, Dong-Won; Woo, Byong-Il; Grace, John Ross
- 발행일
- 2010-01
- 유형
- Article
- 권
- 27
- 호
- 1
- 페이지
- 235 ~ 240