Estimate of the Real-Time Respiratory Simulation System in CyberKnife Image-Guided Radiosurgery

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

1

초록

The purpose of this study was to evaluate the target accuracy according to the movement with respiration of an actual patient in a quantitative way by developing a real-time respiratory simulation system (MISS), including a patient customized 3D moving phantom. The real-time respiratory simulation system (RRSS) consists of two robots in order to implement both the movement of body surfaces and the movement of internal organs caused by respiration. The quantitative evaluation for the 31) movement of the RRSS was performed using a real-time laser displacement sensor for each axis. The average difference in the static movement of the MISS was about 0.01 similar to 0.06 mm. Also, in the evaluation of the dynamic movement by producing a formalized sine wave with the phase of four seconds per cycle, the difference between the measured and the calculated values for each cycle length in the robot that was in charge of body surfaces and the robot that was in charge of the movement of internal tumors showed 0.10 similar to 0.55 seconds, and the correlation coefficients between the calculated and the measured values were 0.998 similar to 0.999. The differences between the maximum and the minimum amplitudes were 0.01 similar to 0.06 mm, and the reproducibility was within +/- 0.5 mm. In the case of the application and non-application of respiration, the target errors were -0.05 similar to 1.05 mm and -0.13 similar to 0.74 mm, respectively, and the entire target errors were 1.30 mm and 0.79 mm, respectively. Based oil the accuracy in the RRSS system, various respiration patterns of patients can be reproduced in real-time. Also, this system dan be used as an optimal too] for applying patient customized accuracy management in image-guided radiosurgery.

키워드

CyberKnifeRRSS (Real-time respiratory simulation system)SynchronyTrackingBREATHING CONTROL ABCORGAN MOTIONTUMOR MOTIONLUNGRADIOTHERAPYIMMOBILIZATIONMOVEMENTPRECISEPATTERNERRORS
제목
Estimate of the Real-Time Respiratory Simulation System in CyberKnife Image-Guided Radiosurgery
저자
Min, Chul KeeChung, Weon KuuLee, SukKim, Chul YongShim, Jang BoHuh, HyunDoLee, SangHoonCho, Sam JuLim, SangwookCho, Kwang HwanLim, Jong SooKwon, Soo Il
DOI
10.3938/jkps.56.164
발행일
2010-01
유형
Article
저널명
Journal of the Korean Physical Society
56
1
페이지
164 ~ 170