Development of a DAQ system for a plasma display panel-based X-ray detector (PXD)
- Lee, Hakjae;
- Jung, Young-Jun;
- Eom, Sangheum;
- Kang, Jungwon;
- Lee, Kisung
WEB OF SCIENCE
0SCOPUS
0초록
Recently, a novel plasma display panel (PDP)-based X-ray detector (PXD) was developed. The goal of this study is to develop a data acquisition system for use with the PXD as an imaging detector. Since the prototype detector does not have any barrier ribs or a switching device in a detector pixel, a novel pixelation scheme the line-scan method is developed for this new detector. To implement line scanning, a multichannel high-voltage switching circuit and a multichannel charge-acquisition circuit are developed. These two circuits are controlled by an FPGA-based digital signal processing board, from which the information about the charge and position of each pixel can be sent to a PC. FPGA-based baseline compensation and switching noise rejection algorithms are used to improve the signal-to-noise ratio (SNR). The characteristic curve of the entire PXD system is acquired, and the correlation coefficients between the X-ray dose, and the signal intensity and the SNR were determined to be approximately 0.99 and 52.9, respectively. (C) 2014 Elsevier B.V. All rights reserved.
키워드
- 제목
- Development of a DAQ system for a plasma display panel-based X-ray detector (PXD)
- 저자
- Lee, Hakjae; Jung, Young-Jun; Eom, Sangheum; Kang, Jungwon; Lee, Kisung
- 발행일
- 2015-06-01
- 유형
- Article; Proceedings Paper
- 권
- 784
- 페이지
- 213 ~ 219
- 언어
- ENG
- 출판사
- ELSEVIER
- 발행국가
- 네덜란드
- 분량
- 7 페이지
- ISSN
- E 1872-9576
P 0168-9002