Pulmonary nodule size evaluation with chest tomosynthesis and CT: a phantom study

  • Shim, S. S.
  • Oh, Y-W
  • Kong, K. A.
  • Ryu, Y. J.
  • Kim, Y.
  • 외 1명
Citations

WEB OF SCIENCE

7
Citations

SCOPUS

8

초록

Objective: We compared digital tomosynthesis (TOMO) and chest CT in terms of assessing the sizes of nodules located in zones where evaluation by simple radiography is limited. Methods: A total of 48 images comprising phantom nodules of four sizes in six different locations were used. Nodule size measurement errors for measurements using TOMO and CT images compared with the actual size from each observer were calculated. The inter- and intra-observer repeatability of the measured values and the agreement between the two techniques were assessed using the method described by Bland and Altman. Results: The mean measurement errors for all of the nodules and four observers were -0.84mm [standard deviation (SD), 0.60mm] on TOMO and -0.18mm (SD, 0.71mm) on CT images. The mean measurement errors for the different observers ranged from -1.11 to -0.55mm for TOMO and from -0.39 to 0.08mm for CT. Assessing the agreement between nodule size measurements using TOMO and CT resulted in mean measurement errors of -0.65mm, with a 95% limit of agreement of -2.53 to 1.22 mm for comparison of TOMO with CT. Conclusion: Our results suggest that nodule sizes obtained using TOMO and chest CT are comparable, even for nodules located in areas where the size measurement is limited on simple radiography. Advances in knowledge: TOMO and CT can be used interchangeably, even for nodules located in a blind area on simple radiography.

키워드

pulmonary noduleCTtomosynthesisRECONSTRUCTIONPARAMETERSOBSERVERLESIONS
제목
Pulmonary nodule size evaluation with chest tomosynthesis and CT: a phantom study
저자
Shim, S. S.Oh, Y-WKong, K. A.Ryu, Y. J.Kim, Y.Jang, D. H.
DOI
10.1259/bjr.20140040
발행일
2015-03
유형
Article
저널명
British Journal of Radiology
88
1047