The reliability of Anycheck device related to healing abutment diameter

  • Lee, Dong-Hoon
  • Shin, Yo-Han
  • Park, Jin-Hong
  • Shim, Ji-Suk
  • Shin, Sang-Wan
  • 외 1명
Citations

WEB OF SCIENCE

18
Citations

SCOPUS

16

초록

PURPOSE. the purpose of this in vitro study was to examine the reliability of the Anycheck device and the effect of the healing abutment diameter on the Anycheck values (implant stability test, IST). MATERIALS AND METHODS. Thirty implants were placed into three artificial bone blocks with 10 Ncm, 15 Ncm, and 35 Ncm insertion torque value (ITV), respectively (n = 10). (1) The implant stability was measured with three different kinds of devices (Periotest M, Osstell ISQ Mentor, and Anycheck). (2) Five different diameters (4.0, 4.5, 4.8, 5.5, and 6.0 mm) of healing abutments of the same height were connected to the implants and the implant stability was measured four times in different directions with Anycheck. The measured mean values were statistically analyzed. RESULTS. The correlation coefficient between the mean implant stability quotient (ISQ) and IST value was 0.981 (P<.01) and the correlation coefficient between the meant periotest value (PTV) and IST value was -0.931 (P<.01). There were no statistically significant differences among the IST values with different healing abutment diameters. CONCLUSION. There was a strong correlation between the Periotest M and Anycheck values and between the ISQ and IST. The diameter of the healing abutment had no effect on the Anycheck values.

키워드

Implant stabilityPeriotestImplant stability testInsertion torque valueImplant stability quotient (ISQ)RESONANCE FREQUENCY-ANALYSISDAMPING CAPACITY ASSESSMENTIMPLANT STABILITY
제목
The reliability of Anycheck device related to healing abutment diameter
저자
Lee, Dong-HoonShin, Yo-HanPark, Jin-HongShim, Ji-SukShin, Sang-WanLee, Jeong-Yol
DOI
10.4047/jap.2020.12.2.83
발행일
2020-04
유형
Article
저널명
Journal of Advanced Prosthodontics
12
2
페이지
83 ~ 88