Development of a system for on-site dosimetry audit of high-dose-rate brachytherapy

  • Gwon, Hyeok-Jun; 
  • Lee, Wonho; 
  • Choi, Sang-Hyoun; 
  • Cho, Gyu-Seok; 
  • Lee, Soon Sung; 
  • 외 2명
Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

Despite international standards, the lack of an independent audit system in South Korea has led to quality management challenges in high-dose-rate brachytherapy, prompting the development of a precise source activity measurement system and a regulatory MATLAB code. We developed an audit system that integrates a source strength measurement system and mechanical accuracy check phantoms using a well-type chamber electrometer setup and an in-house MATLAB code. We achieved audit suitability with appropriate uncertainty and compatibility across various Remote After-Loading System (RALS). Our uncertainty assessments revealed the proposed expanded uncertainties (k = 2) of 2.72% and 2.71% for models 70010 and 72280, respectively. Based on these results, we plan to implement this auditing system nationwide to comply with domestic regulations.

키워드

HDR brachytherapy; On-site dosimetry audit; ISO/IEC17025; Audit procedure; HDR; CHAMBER
제목
Development of a system for on-site dosimetry audit of high-dose-rate brachytherapy
저자
Gwon, Hyeok-Jun; Lee, Wonho; Choi, Sang-Hyoun; Cho, Gyu-Seok; Lee, Soon Sung; Hwang, Sun-Boong; Kim, Kum Bae
DOI
10.1007/s40042-025-01331-8
발행일
2025-03-10
유형
Article; Early Access
저널명
Journal of the Korean Physical Society
권
86
호
9
페이지
909 ~ 917